Hach AutoCAT 9000 Instruction Manual page 77

Chlorine amperometric titrator
Table of Contents

Advertisement

9.
The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
to change one
CHECK MARK
of the analysis parameters or
press 1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if autoscaling software is
enabled.
50081_ChlorineDioxideGeneratorYield.fm
10.
The Start Timer will
begin, allowing the
electrode to stabilize before
data is collected.
After the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve in real
time.
Chlorine Dioxide Generator Yield
11.
The Zoom range
determination window will
display the titration curve.
Select and move the cursors
to define the range of data
points for analysis.
Proceed to step 12.
Chlorine Dioxide Generator Yield
12.
Select type of end point
determination.
Press
for Manual.
DEL
Note: Automatic determination
attempts to identify linear
segments without user input
but is not recommended for
this method.