Specification Tables

Specification
Digital indicators and Display
Display Update Rate
Status Display
Communications
Mathematics
Front and Rear USB Ports
Standard Screens and Custom
Screens
43-TV-25-35 GLO Iss.4 Dec 06 UK
A5E01001767-04
Specification Tables
D300 Recorder
Display size and Type: 5.5" diagonal, Digital Colour LCD (TFT) with Touch Screen
Industrial grade with brightness adjustment and wide viewing angle
Resolution: QVGA (320 x 240 pixels).
Screen Saver: Set in minutes from 1 to 720, can be set to dim the screen or to switch off.
Brightness adjustment: Adjustable between 10 and 100%, default set to 80% brightness.
Backlight life time: 55,000 hours to half brightness when used at 100%, (86,000hr if used at 80%).
Maximum luminosity 400 cd/m
Touch Screen life: 1,000,000 touches
D400 Recorder
Display size and Type: 12.1" diagonal, Digital Colour LCD (TFT) with Touch Screen
Industrial grade with brightness adjustment and wide viewing angle
Resolution: SVGA (800 x 600 pixels).
Screen Saver: Set in minutes from 1 to 720, can be set to dim the screen or to switch off.
Brightness adjustment: Adjustable between 10 and 100%, default set to 80% brightness.
Backlight life time: 43,000 hours to half brightness when used at 100%, (67,000hr if used at 80%).
Maximum luminosity 400 cd/m
Touch Screen life: 1,000,000 touches
D200 Recorder
Display size and Type: 5" diagonal, Digital Colour LCD (TFT) with Touch Screen
Industrial grade with brightness adjustment and wide viewing angle
Resolution: QVGA (320 x 240 pixels).
Screen Saver: Set in minutes from 1 to 720, can be set to dim the screen or to switch off.
Brightness adjustment: Adjustable between 10 and 100%, default set to 80% brightness.
Backlight life time: 40,000 hours to half brightness when used at 100%, (62,500hr if used at 80%).
Maximum luminosity 450 cd/m
Touch Screen life: 1,000,000 touches
Display values updated every second
A status bar, at the top of the recorder's screen, displays the real-time icons of the recorder status,
such as Time left and alarm active.
Ethernet 10/100 base - T with RJ45 connector supporting Modbus/TCP, FTP, Internet, DHCP or
fixed IP address. RS485Modbus RTU (up to 115200 Baud Rate). RS485 is available as an option
on the Comms card for the D200.
Basic maths include Add, Subtract, Multiply, Divide, Modulo and power. Full Maths and Scripting
(option) support up to 100 character free form math expression for each pen. Like SINE, COS,
TAN, Log, Parenthesis (eg. A1 + A2), comm variables, free memory, and access to any data item
variable (A1, P1, D1 etc.). (Scripting is not available for the D200).
USB host ports front and rear for data and setup transfers through these ports. External devices
keyboard or mouse, Barcode reader, or external mass storage device. (Rear USB Host is available
as an option on the Comms card for the D200)
Fully programmable display values in engineering units. Time & date stamp on every division, pro-
viding the current time and date.
Sets of Standard screens are available to display data on a chart, digital reading, bargraphs or
numerous combinations thereof. Screen properties can be modified on the recorder and custom-
ised to suit. Custom screens created in the Screen Designer software can be imported into the
recorder for specialist applications. Custom Screen firmware option is required.
Digital values displayed include alarms on bars, engineering units, pen name, tag, time and date,
20 character description and totalised values. (Custom Screens are not available as an option for
the D200)