(Main Menu > Configure > Setup > Edit Setup> General > Factory > FW Upgrade
The Firmware Upgrade button allows new versions of the recorder firmware to be loaded
into the recorder via a USB key or a Compact Flash card. The file type required to upgrade
the Firmware has the file extension .xsu. Contact Honeywell for more information.
Firmware downloads are available from
Insert the Compact Flash or USB key with the latest revision of Firmware (.xsu file format)
and press the FW Upgrade button. The recorder will scan and check the files on the external
device with the current version running in the recorder. Any later version files on the external
device are copied into the recorder.
If there is no external media fitted or the file is corrupt a Firmware installation dialog will ap-
pear. If this occurs check the file is the correct format (.xsu). Secondly download the file to
another Compact flash card or USB key.
See "Storage Media Format" on page 146
cards and USB keys. Compact Flash not available
To check the firmware upgrade has been successful, see the Status Menu,
page 156.
(Configure > Setup > Edit Setup> General > Factory > Calibration)
Cards will be calibrated in the factory. All ranges are factory calibrated and set to the default
range of +/-12V.
Date & Type of last calibration is stored in
Analogue In and CJC Calibration can be performed per slot (card). Select the AI or CJC cal-
ibration button to go to the next menu.
In this menu there is a button for each Analogue Input card fitted in the recorder. Calibration
is done per channel.
Cards may be reset to original 'factory' cal, i.e. a card holds two sets of calibration constants
(factory & user) for all ranges.
Factory Cal - The recorder is calibrated in the factory. These are pre-stored calibration val-
ues that cannot be changed by the user. If the recorder is re-calibrated to a User Cal the
factory values can be restored at any time.
User Cal - User calibration can be applied at any time. To meet the accuracy values spec-
ified (see
source should be accurate to 0.01% full scale or better.
On initial power up the Factory Cal and the User Cal are the same.
43-TV-25-30 Iss.6 GLO Aug 07 UK
• Options Code - The Options code is a unique coded number for this recorder that con-
tains information that enables certain functionality including the total credit value.
• Options - This displays how many credits are currently in use. Select this to produce a
list of Firmware options available and their credit values. Toggle each option On or Off
to enable the feature within the recorder. If there are not enough credits available. More
credits are available from your supplier. For a complete lists of all the firmware options
and their functionality see

FW Upgrade

Calibration

Table 4.7, "Calibration Input range table," on page

AI Calibration

"Input Range Performance Accuracy Table" on page 263.),
Table 7.1, "Firmware Options," on page 201
http://hpsweb.honeywell.com
for formatting information of Compact Flash
for the
"Maintenance" on page 159.
Table 4.7, "Calibration Input range table," on page 105.
eZtrend QXe
.
"General" on
105.
calibration
103