Continuous M-Mode Measurements - GE LOGIQ 500 Advanced Reference Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

Advanced Cardiac Calculations

Continuous M-Mode Measurements

RVDd
IVSd
LVIDd
LVPWd
Figure 4–16. Continuous M-Mode Measurements
4–42
When M-Mode measurements can be taken on the same
timeline, the system will support a continuous measurement of
all values along that line.
This can be done because the end point of one measurement is
the start point of the next measurement. After the end point of
the last measurement is set, keystrokes are reduced by not
having to set the next start point.
This type of M-Mode measurements is supported by M-LV/RV,
Teichholz, Cubed and Gibson menu selections.
After selecting the measurement sequence, use the Trackball
to position the first cursor and press Set. The name of the first
value appears; use the Trackball to position the cursor at the
end of the first measurement and press Set. This is the start
point of the next measurement. The name of the next
measurement appears; use the Trackball to position the end
point of the second measurement and press Set.
This process continues until all measurements are taken.
RV
IVS
LV
LVPW
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0