FDA Acoustic Output Data
FDA Acoustic Output Tables
Maximum output summary
6–18
The following tables list the typical maximum acoustic output
levels achievable with the LOGIQt 500 for all probes and
operational modes. It is intended that this information be useful
in making ALARA decisions and selecting the most appropriate
probe for the application. In accordance with US FDA
Guidelines, the overall maximum acoustic SPTA intensity for
LOGIQt 500 is limited to 720 mW/cm
Modes for which TI does not exceed 0.1 are indicated by <0.1.
2
and MI is limited to 1.9.
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0