Exchanging Heat Exchanger Fan

When required
Preparations for this
procedure
Parts required
1 Unplug the connector at the TCA board and remove the
wire from the Top Foam.
3 Remove the Fan from the Fan Gasket.
1100 Series AT Reference Manual
When defective
Removing the Top Cover and Foam (see
on page 63) item 1 to 4.
Fan, PN 3160-1079
2 Carefully lift the Fan upwards.
4 Insert the new Fan in the Fan Gasket. See arrow at the side
of Fan for correct airflow direction.
Repairing the ALS Thermostat Module
"Removing the Top Cover and Foam"