Controller Requirements

Agilent 1100 Series control module and Agilent ChemStation must have the
latest revision firmware/software loaded to allow optimum operation with the
thermostattable autosamplers. Older revisions might not recognize or have full
functionality with the thermostattable autosampler.
Control Module Firmware requirements
The control module requires firmware revision A.01.30 or higher to control
the thermostatted autosampler. Previous firmware revisions of the control
module will not run with the thermostatted autosampler. If the control module
was shipped together with the thermostatted autosampler the control module
firmware does not require updating. The firmware update must be done with
a PCMCIA card that has the newest revision loaded. The firmware is not part
of the shipment of the thermostatted autosampler. Contact your local Agilent
Technologies sales and service office for the firmware update of the control
module. For the update procedure see
Firmware"
Agilent ChemStation Software requirements
To control the thermostatted autosampler from a PC, the Agilent ChemStation
software version A.04.03, or A.05.02beta, or A.05.02 or higher is required.
These software revision might however not support all ALS modules. The
thermostatted autosampler will not run with any previous version of the
Agilent ChemStation software. The software updates are part of the
thermostatted autosampler shipment. For the update procedure see
Update of the Agilent ChemStation Software"
1100 Series AT Reference Manual
on page 40.
Installing the G1330B Thermostat
"Stage 7: Update of Control Module
on page 41.
2
"Stage 8:
27