3
Troubleshooting and Test Functions

Temperature Sensor Failed (1 - 4)

The ALS thermostat is equipped with four sensors to monitor the performance
of the instrument. The sensors are connected to both sides of the Peltier
elements for control of the cooling / heating efficiency. As the positioning of
the sensor is critical to meet the performance requirements of the ALS
thermostat the single sensors cannot be replaced separately.
Probable Causes:
• Temperature sensor is defective
Suggested Action:
52
See
""Exchanging the Sensors"
on page 69".