Installing the G1330B Thermostat

Stage 6: Turning on the Thermostatted Autosampler

1 Depress the power switch to turn on the two modules of the autosampler.
The power switch stays depressed (1) and a green indicator lamp in the power switch is on
NO TE
when the autosampler is turned on. When the line power switch stands out (Ø) and the
green light is off, the autosampler is turned off.

Stage 7: Update of Control Module Firmware

If the control module has a firmware revision A.01.30 or higher, you do not
need to update the control module firmware.
If you have a control module version A.01.30 or lower, update the firmware as
described.
1 Disconnect the control module before inserting the PC card.
2 Insert the PC card into the card slot of the control module.
3 Reconnect the control module for restarting it.
4 Press "System" (F5) - "Records" (F4). Highlight the LC-System line in the
5 Press "FW-Update" (F5).
6 Select the file for the firmware update (LCB202en.BIN).
7 Press "Execute" and select "Yes" to confirm loading of the new firmware.
8 Check that the correct firmware was loaded by pressing "System" (F5) -
9 Disconnect the control module and remove the PC card by pressing the
40
display using the up-down arrows.
The control module reboots and loads the firmware indicated by (.) and (*)
on the display. When finished with the update the control module reboots
again.
"Records" (F4).
card-eject button.
1100 Series AT Reference Manual