Removing the Top Cover and Foam

Tools required
Preparations for this
procedure
Do not disconnect or reconnect the autosampler to ALS thermostat cable when the
WA RN ING
power cords are connected to either of the two modules. This will damage the
electronics of the modules.
1 Remove the front cover by pressing the both clip fasteners
on both sides of the cover.
1100 Series AT Reference Manual
Screwdriver Pozidriv #1
Switch off ALS at the main power switch. Disconnect ALS and ALS thermostat
power cords. Remove ALS to ALS thermostat cable, and remove thermostat
from stack.
2 Move the lever towards the power socket.
Repairing the ALS Thermostat Module
4