3.3.7 Trimming the Outputs

The Trim Menu enables the operator to compensate for differences in
measurement of the 0/4-20 mA or 0-2V outputs by connected recorders or
SCADA equipment. To trim the output:
This will cause the MTS6 to output 4.000 mA or 0.4 V on the output being
trimmed. The output value should then be read using the connected
recorder, SCADA equipment, or DVM. Enter the value read from the
connected equipment as the Zero Trim value, as follows:
Since you cannot trim 0 mA or 0 V for negative offsets, trim for the
Note:
lower end of the scale is at the 4 mA or 0.4 V output level.
Moisture Target Series 6 Hygrometer User's Manual
Chapter 3. Operation and Programming
Select Trim from the Output Menu and press
Enter
. The following screen appears.
When performing a Trim operation, the
MTS6 unit requires you to first reset the trim.
To reset the trim output, select Reset Trim
and press Enter
appears.
Use the left or right arrow keys to select YES
and press Enter
previous trim values, and returns the MTS6
to its factory adjustment. The display returns
to the previous screen.
To trim the zero value, select Trim Zero and
press Enter
. A screen similar to the
following appears.
. The following screen
. This cancels any
41