Using Setup Load - Fluke Hydra Series II User Manual

Data bucket
Table of Contents

Advertisement

2635A
Users Manual

Using Setup Load

Perform the procedure in Figure 3-6 to discard the current instrument configuration and
load a configuration saved in a previous SETUP STORE operation (Figure 3-5). A
configuration file includes channel configurations, scan interval, measurement rate,
alarms, Mx+B scaling, and temperature unit (ºC or ºF). Communication parameters,
baud, parity, CTS, and echo remain as set previously by the front panel controls. To exit
at any time (file not loaded), press the C key.
FILES
ENTER
ENTER
ENTER
3-10
FILES
Init
StAt
dir
dAtA
SEtUP
SEtUP
ErASE
StorE
LOAd
LoAd
SEt99
SEtxx
SEt00
Typical
Required
Figure 3-6. Using SETUP LOAD to Load Configuration Files
Selecting the SETUP mode. Press the FILES
key to access the FILES menu.
up/down arrow keys until SEtUP is displayed,
then press the ENTER key. The menu changes
to SEtUP.
If an error message appears, see
Table 3-1.
Selecting the LOAD mode. Press the up/down
arrow keys until LoAd is displayed in the SETUP
menu, then press the ENTER key. The menu
changes to LoAd.
Selecting a SEtxx File to LOAD.
up/down arrow keys until the SEtxx file to be
loaded is displayed in the LoAd menu, where xx
represents a number from 00 to 99, then press
the ENTER key. After ENTER is pressed, the
entire meter configuration changes to reflect the
data stored in the selected SETUP file. A display
of nonE indicates no setup files are stored.
Press the
Press the
op42f.eps