2635A
Users Manual
ALARM_ASSOC?
ALARM_ASSOC_CLR
4-28
Table 4-5. Command and Query Reference (cont)
Alarm Association Query
Returns alarm output associations at the rear panel DIGITAL I/O connector
for channels 4 to 20.
ALARM_ASSOC?
<channel>,<limit_num>
<channel>
= (4 .. 20)
<limit_num> = 1 or 2
This command returns an integer that represents the DIGITAL I/O line
active at the rear panel DIGITAL I/O connector for the specified channel
and alarm limit. If default settings are in effect, returns follow the table
below.
I/O 4
I/O 5
Chan
Chan
4
5
8
9
12
13
16
17
20
If there is no association between an alarm and DIGITAL I/O line, there is
no return and an Execution Error is generated.
Example: ALARM_ASSOC? 10,1 returns 2 [For channel 10, alarm Limit 1,
will assert DIGITAL I/O line 2].
Clear Alarm Association
Clears an alarm output association at the rear panel DIGITAL I/O
connector for channels 4 to 20.
ALARM_ASSOC_CLR <channel>,<limit_num>
This command removes all association between a DIGITAL I/O line at the
rear panel DIGITAL I/O connector for the specified channels 4 to 20 and
alarm limit. After application of this command, the previously associated
DIGITAL I/O line is set high and new alarm conditions on this channel's
alarm limit will not assert the DIGITAL I/O line. If this command is entered
during scanning while logging to the memory card, and Execution Error is
generated.
Example: ALARM_ASSOC>CLR 10,1 [Channel 10, alarm Limit 1, remove
all association with a DIGITAL I/O line.]
I/O 6
I/O 7
Chan
Chan
6
7
10
11
14
15
18
19
<channel>
= (4 .. 20)
<limit_num> = 1 2