Fluke Hydra Series II User Manual page 155

Data bucket
Table of Contents

Advertisement

SCALE_MB
Set Mx+B Scaling Values
Set the M and B scaling values for the indicated channel, and display the results of
the Mx+B calculation in the indicated display range. Changing the Mx+B of any
channel also clears the Review array, and resets ALARM OUTPUTS and DIGITAL
I/O lines.
SCALE_MB
The range code for the display <range> is shown below
Range
Code
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
When M=1 and B=0, there is no Mx+B scaling. The entries for M and B must be
between +/-0.0001E-3 and +/-9999.9E+6. An Execution Error is generated by invalid
entries, a channel set to OFF, if the instrument is
scanning, or if the range code is too low for the selected B value. For example, the
minimum display range for B=1000 is code 8. Mx+B scaling values for a channel are
automatically reset to 1 (M) and 0 (B) when the function for that channel is changed.
Returned measurements for a channel with Mx+B scaling has a function identifier of
MX+B (when FORMAT 2 has been asserted).
Example: SCALE_MB 18,+.55555,-17.777,6 [For channel 18, M=+.55555, B=-
17.777, and the display range is 00.000 x1.]
Table 4-5. Command and Query Reference (cont)
<channel>,<M_value>,<B_value>,<range>
<channel> = (0 .. 20)
<M_value> = signed numeric quantity
<B_value> = signed numeric quantity
<range> = (1 .. 16)
Display
Max B
Offset
Value
0.0000 m
9.9999E-3
00.000 m
99.999E-3
000.00 m
999.99E-3
0000.0 m
9999.9E-3
0.0000 x1
9.9999
00.000 x1
99.99
000.00 x1
999.99
0000.0 x1
9999.9
0.0000k
9.9999E3
00.000k
99.999E3
000.00 k
999.99E3
0000.0 k
9999.9E3
0.0000 M
9.9999E6
00.000 M
99.999E6
000.00 M
999.99E6
0000.0 M
9999.9E6
Computer Operations
Computer Interface Commands and Operation
4
4-51