Setting the Scan Interval

Perform the procedure in Figure 2-11 to set the time between starts of measurement
scans. In preparation, the instrument must be in the inactive mode (not scanning or
monitoring). To exit at any time (changes not saved), press the C key. The scanning
interval format is HOURS:MINUTES:SECONDS. The minimum is 0:00:00 (continuous
scanning [default]); the maximum is 9:99:99 (9 hours, 99 minutes, 99 seconds). The scan
interval is divided into two portions: the measurement interval when measurements are
actually taken, and the time-out interval that completes the overall scan duration. For
example, if 10 channels can be measured in 8 seconds, and the scanning interval is set
for 30 seconds, the first 8 seconds are used for measurement, while the remaining 22
seconds are used to time out. If the scanning interval is set to less than the measurement
rate, the effect is continuous scanning. For example, if 10 channels can be measured in 8
seconds and the scanning interval is set for 5 seconds, scanning is continuous. To speed
up the measurement rate, refer to Figure 2-12.
INTVL
ENTER

Setting the Scan Interval.

SET
(Interval) key to access the scan time menu. The
0: 0 0 : 0 0
format is HOURS:MINUTES:SECONDS.
- to -
the up/down and left/right arrow keys to select
9: 9 9 : 9 9
and configure each column in the menu.
example, a scan interval of 1 hour, 25 minutes,
and 33 seconds would be formatted as 1:25:33.
The column being configured will have a bright
Typical
display.
Required
interval, press the ENTER key.
Figure 2-11. Setting the Scan Interval
Front Panel Operations
Setting Operating Conditions
Press the INTVL
When the display shows the desired
2
Press
For
op23f.eps
2-17