726
Users Manual
Manually Stepping the mA Output
To manually step current output:
•
Use
or
H
I
25 % steps.
•
Touch either
J
100 %.
Auto Ramping the Output
Auto ramping can continuously apply a varying stimulus
from the Calibrator to a transmitter, while your hands
remain free to test the response of the transmitter.
L is pressed
When
repeating 0 % - 100 % - 0 % ramp in a choice of three
ramp waveforms:
E
•
0 % - 100 % - 0 % 40-second smooth ramp
P
•
0 % - 100 % - 0 % 15-second smooth ramp
N
•
0 % - 100 % - 0 % Stair-step ramp in 25 %
steps, pausing 5 seconds at each step. Steps are
listed in Table 7.
42
information@ITM.com
to step the current up or down in
to go to 0 %, or
to go to
G
, the Calibrator produces a
Fluke-Direct
To exit ramping, press any button.
Table 7. mA Step Values
Step
0 %
25 %
50 %
75 %
100 %
Storing and Recalling Setups
Up to eight settings can be stored in a nonvolatile
memory to recall for later use. A low battery condition or
a battery change does not jeopardize the stored settings.
Store a Setup
To store a setup:
1.
Create the desired setup.
2.
Push S. The right side of the display changes to
show SAVE SETUP and SAVE DATA.
3.
Push Sto select SAVE SETUP.
4.
Push Y or Z to select the desired memory location (at
the top of the LCD).
5.
Push S to enter the setup.