Monitoring
The RTR-57C can be set to patrol and monitor RTR-51/52/53 or
RVR-52 groups and display in order the current readings for view.
1.
In the Main Menu, open [ MONITOR MODE].
2.
Select the monitoring range.
[All Groups] :
Remote Units in all groups registered to the RTR-57C will be contacted
via wireless communication, and current measurement readings for those
units with which communication is possible will be displayed consecutively
every 2 seconds. This can be done for up to 120 units; the 121
reading cannot be displayed.
[Specify Group] :
All Remote Units in the selected groups will be contacted via wireless
communication, and current measurement readings for those units will be
displayed consecutively every 2 seconds.
For those units with which communication is impossible the measurement
will be displayed as [----].
3.
Select and set the Monitoring Interval
[Interval] :
The monitoring interval is the amount of time between the starting of
wireless communication to gather current Remote Unit readings. Selections
can be made from 15 sec, 30 sec, and 1 to 60 minutes (in 1 minute units).
NOTE:
The amount of communication time depends on the number of registered
●
groups and remote units. Please note that if the monitoring interval is set
for less time than communication takes to complete a round of monitoring,
the system will automatically extend the interval to allow for the necessary
amount of communication time.
47
NOTE:
The shorter the monitoring interval the sooner batteries will loose their
●
power.
<Communication Time for [All Groups] >
●
・If there are 64 Remote Units registered to the group
(24 seconds per group x Number of groups)
・If there are 250 Remote Units registered to the group
(54 seconds per group x Number of groups)
Moreover, the display time for each unit is about 2 seconds.
<Communication Time for [Specify Groups]>
●
unit's
st
The communication time will depend on the number of registered remote
units. For 250 units registered to 1 group the time will be 54 seconds plus a
display time for each unit of 2 seconds.
4.
By selecting the monitoring range and clicking, the monitoring
search will begin.
If using [Specify Group], by specifying the group and clicking, the
●
monitoring search will begin.
5.
After the search has been completed, the current measurement
readings for those remotes registered in the group(s) will be
consecutively displayed.
For RTR-53, the display will show in the order of temp / humidity.
●
If the buzzer has been set to ON and a result is NG, a beeping sound will
●
go off.
【If the upper and lower limit
settings are activated.】
Result
Upper and Lower Limits
48