Wireless Communication:
Wireless Communication:
The following settings and changes can be made from the main unit:
Recording Start Date and Time (Programmed Start)
Recording Mode
Recording Interval
※ To change the unit of temperature ( ゜ C or ゜ F) in the RTR-51/52/53 units and the
measurement mode in the RVR-52 units, please use your computer and the
software『T&D Recorder for Windows 』 .
NOTE:
If you make recording condition changes while recording, all previously
recorded data will be lost. Please make sure to download all recorded data
before making changes.
Changes in recording conditions will result in a lifting of the upper and lower
limits that had been set for each unit. If you wish for those settings to remain
in effect, please make sure to use the [Read Settings] function before
making any changes. (See p.24 for details.)
【Immediate Start】
1.
In the Main Menu, click [REC SETTINGS] and then [ Wireless].
2.
Select the Remote Unit for which you wish to make settings.
① Select a Group in which
Remote Units are registered.
19

Recording Start

Recording Start
② Select a Remote Unit or a
Group.
Before selecting a Remote Unit you can run a search to find all Remote
Units with which communication is possible. Those units with which
communication is possible will appear with a
(See p.45 for details)
3.
Make Recording Condition Settings.
If you wish to make no changes and want to restart recording under the
same conditions, please read the current settings first and then restart
recording. (See p.24 for details)
Recording Modes:
One Time: Recording will stop when the data capacity becomes full.
Endless: Recording will continue with oldest data continuously
overwritten.
Recording Interval:
Choose from 15 intervals: 1,2,5,10,15,20,30 seconds and 1,2,5,10,15,20,
30 and 60 minutes.
4.
After having completed the settings, adjust the arrow (>) to [REC
Immediate] and press dial to transmit the recording conditions to the
Remote Unit.
mark next to them.