Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact
ResMed.

Alarm troubleshooting

The most common reason for an alarm to sound is because the system has not been properly
assembled. Check that the air tubing has been properly attached to the device and patient
interface (and humidifier if used).
Notes:
The alarm actions listed below are based on having the appropriate alarm settings for the
patient's therapy. When a user adjustable alarm is activated, re-confirm the alarm settings.
The alarm log and alarm settings are maintained when the device is powered down and in the
event of a power loss.
If multiple alarms are active simultaneously, the alarm with the highest priority will be
displayed first.
If an alarm activates repeatedly, discontinue use and return the device for servicing.
Problem/Possible cause
LCD: Internal Battery Empty!
The remaining battery charge is below 15%.The
device can be powered by the internal battery for
maximum 2 minutes.
System Failure!
LCD:
Component failure.
The device stops delivering air pressure (system
failure 6, 7 , 9, 22, 38).
Therapy cannot be started (system failure 21).
Component failure (system failure 8, 25).
The following system failures can occur: motor failure (system failure 6), pressure sensor failure
(system failure 7), software failure (system failure 8), calibration fault (system failure 9), self test failure
(system failure 21), flow sensor failure (system failure 22), system component failure (system failure
25), critical software failure (system failure 38).
LCD: Over Pressure!
The device generates a pressure that is greater
than 60 cm H
LCD: Blocked Tube!
Air path is blocked.
O. Treatment will be stopped.
2
Action
Connect the device to mains power.
Note: In case of a total power failure, the therapy
settings will be stored and therapy will resume
when the device is powered again.
1. Power-off the device.
2. Power-on the device again.
If the problem persists return the device for
servicing.
1. Power-off the device, if the system failure is
continuously present.
2. Power-on the device again.
If the problem persists return the device for
servicing.
1. Power-off the device.
2. Check that the air tubing is connected properly.
3. Power-on the device again.
4. Start Learn Circuit function.
Note: If the alarm activates repeatedly internal
components may be defective. Discontinue use
and return the device for servicing.
1. Check the air path for any blockages.
2. Remove the blockages.
3. If the alarm is not cleared, stop treatment.
4. Re-start treatment.