Alarms
2. Press Reset all to reset multiple alarms. Only those alarms that can be reset, will be reset. Any
remaining alarms will require user intervention and correction.
3. Complete any required action to resolve the remaining alarms.
4. Press OK to close the Active alarms screen and return to the previous screen.

Adjusting alarm settings

CAUTION
Adjusting alarm thresholds to maximum or minimum values may render the alarm ineffective.
WARNING
The safety and effectiveness of alarms settings should be verified for each enabled Program.
Note: To adjust the volume of the audible alert, refer to Device settings (see page 31).
To adjust the alarm settings for the current active program:
1. Access Clinical mode. The Settings screen is displayed.
2. Press
.
3. Press the alarm setting that you want to adjust. The selected setting is highlighted in orange and the
up and down scroll arrows appear. Minimum and maximum settable values appear with the scroll
arrows.
4. Increase or decrease the alarm setting as required.
5. To adjust other alarm settings, select the desired settings and repeat this process. All settings that
have been adjusted are displayed with an orange outline.
6. When no further adjustments are required, press Apply.
Note: If a change to the High Pressure alarm setting conflicts with one or more ventilation parameter settings, a
confirmation screen is displayed providing a summary of the revised ventilation parameter settings. To accept
these settings press Confirm.