Adding each instrument to the network using BootP
1 Follow Start > All Programs > Agilent BootP Service and select Edit BootP
2 Uncheck the Do you want to log BootP requests? once all instruments have
3 Click Edit BootP Addresses... The Edit BootP Addresses screen appears.
4 Click Add... The Add BootP Entry screen appears.
5 Make these entries for the instrument:
1290 Infinity Quaternary Pump User Manual
Settings. The BootP Settings screen appears.
been added.
The Do you want to log BootP requests? box must be unchecked when you
have finished configuring instruments; otherwise, the log file will
quickly fill up disk space.
Figure 47
Enable BootP logging
• MAC address
• Host name, Enter a Hostname of your choice.
The Host Name must begin with "alpha" characters (i.e. LC1260)
• IP address
• Comment (optional)
LAN Configuration
Automatic configuration with Bootp
13
277