9132 Portable IR Calibrator
User's Guide
7.4
7.4.1
7.4.2
7.5
24

Scan

The scan rate can be set and enabled so that when the set-point is changed the
dry-well heats or cools at a specified rate (degrees per minute) until it reaches
the new set-point. With the scan disabled the dry-well heats or cools at the
maximum possible rate.

Scan Control

The scan is controlled with the scan on/off function that appears in the main
menu after the temperature scale units.
Scan function off
Sc=OFF
Press "UP" or "DOWN" to toggle the scan on or off.
Scan function on
Sc=On
Press "SET" to accept the present setting and continue.
S
Accept scan setting

Scan Rate

The next function in the main menu is the scan rate. The scan rate can be set
from .1 to 99.9°C/min. The maximum scan rate however is actually limited by
the natural heating or cooling rate of the instrument. This is often less than
100°C/min, especially when cooling.
The scan rate function appears in the main menu after the scan control function.
The scan rate units are in degrees per minute, degrees C or F depending on the
selected units.
Sr=10.0
Scan rate in C/min
Press "UP" or "DOWN" to change the scan rate.
New scan rate
Sr=2.0
Press "SET" to accept the new scan rate and continue.
S
Accept scan rate

Set-point Resistance

To display the Set-point Resistance, press the "SET" and "DOWN" keys simul-
taneously when the temperature is displayed. When the "SET" and "DOWN"