Scan - Fluke 9103 User Manual

Dry-well
Hide thumbs Also See for 9103:
Table of Contents

Advertisement

9103 Dry-Well
User's Guide
7.3
7.3.1
7.3.2
7.4
22
info@Fluke-Direct.com

Scan

The scan rate can be set and enabled so that when the set-point is changed the
calibrator heats or cools at a specified rate (degrees per minute) until it reaches
the new set-point. With the scan disabled the calibrator heats or cools at the
maximum possible rate.
Scan Control
The scan is controlled with the scan on/off function that appears in the main
menu after the set-point function.
Sc=OFF
Scan function off
Press "UP" or "DOWN" to toggle the scan on or off.
Sc=On
Scan function on
Press "SET" to accept the present setting and continue.
S
Accept scan setting
Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set
from .1 to 99.9°C/minute. The maximum scan rate however is actually limited
by the natural heating or cooling rate of the instrument. This rate is often less
than 100°C/minute, especially when cooling.
The scan rate function appears in the main menu after the scan control function.
The scan rate units are in degrees per minute, degrees C or F depending on the
selected units.
Scan rate in °C/min.
Sr=10.0
Press "UP" or "DOWN" to change the scan rate.
Sr= 2.0
New scan rate
Press "SET" to accept the new scan rate and continue.
S
Accept scan rate
Temperature Display Hold
The calibrator has a display hold function which allows action of an external
switch to freeze the displayed temperature and stop the set-point from scan-
ning. This function is useful for testing thermal switches and cutouts. This sec-
Fluke-Direct
.com
1.888.475.5235