User's Guide

Scan

Set and enable the scan rate so that when the set-point is changed the calibrator heats or cools at a
specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled, the
calibrator heats or cools at the maximum possible rate.

Scan Control

The scan is controlled with the scan on/off function that appears in the main menu after the set-point
function.
Sc=OFF
Push UP or DOWN to toggle the scan on or off.
Sc=On
Push SET to accept the present setting and continue.
SET

Scan Rate

The scan rate can be set from .1 to 99.9 °C/minute. The maximum scan rate is actually limited by the
natural heating or cooling rate of the Product. This rate is often <100 °C/minute, especially when
cooling.
The scan rate function appears in the main menu after the scan control function. The scan rate units
are in degrees C or F per minute, depending on the selected units.
Sr=10.0
Push UP or DOWN to toggle the scan on or off.
Sr=2.0
Push SET to accept the new scan rate and continue.
SET

Temperature Display Hold

The Product has a display hold function which allows an external switch to freeze the displayed
temperature and stop the set-point from scanning. Use this function to test thermal switches and
cutouts. This section explains the functions available to operate the temperature hold feature.
A subsequent example shows how to set up and use the hold feature to test a switch.
Scan function off
Scan function on
Accept scan setting
Scan rate in °C/min
New scan rate
Accept scan rate