9132 Portable IR Calibrator
User's Guide
9.2.2
36
=
−
B
T
T
2
1
⎡
⎤
⎡
⎤
T
T
=
−
−
3
3
C
1
⎣ ⎢
⎦ ⎥
⎣ ⎢
⎦ ⎥
100
100
⎡
⎤
⎡
⎤
T
T
=
−
−
2
2
D
1
⎣ ⎢
⎦ ⎥
⎣ ⎢
⎦ ⎥
100
100
=
−
E
R
R
3
2
=
−
F
R
R
2
1
−
AF
BE
=
delta
−
DE CF
T
- Measured temperature using a thermometer
1-3
R
– Value of the set-point resistance from the display. (Press SET and
1-3
DOWN at the same time.)
Where:
T
and R
are the measured temperature and resistance at 50.0°C
1
1
and R
are the measured temperature and resistance at 200.0°C
T
2
2
T
and R
are the measured temperature and resistance at 350.0°C
3
3
Compute R0 and ALPHA
⎡
⎤
⎡
T
=
+
1
a
T
delta
⎣ ⎢
⎦ ⎥
⎣ ⎢
1
1
100
⎡
⎤
⎡
T
=
+
3
a
T
delta
⎣ ⎢
⎦ ⎥
⎣ ⎢
3
3
100
−
R a
R a
=
3 1
1 3
rzero
−
a
a
1
3
−
R
R
=
1
3
alpha
−
R a
R a
3 1
1 3
Where:
delta is the new value of DELTA computed above
Program the new values for DELTA (delta), R0 (rzero) & ALPHA (alpha) into
the instrument using the following steps.
1. Press the SET and EXIT keys at the same time and then press SET until
R0 is displayed.
⎡
⎤
⎡
⎤
T
T
−
2
2
1
⎣ ⎢
⎦ ⎥
⎣ ⎢
100 ⎥
⎦
100
⎡
⎤
⎡
⎤
T
T
−
1
1
1
⎣ ⎢
⎦ ⎥
⎣ ⎢
100 ⎥
⎦
100
⎤
T
−
1
1
⎦ ⎥
100
⎤
T
−
3
1
⎦ ⎥
100