Adjust the setting using "UP" or "DOWN" and pressing "SET".
d=HALF
7.12.4

Linefeed

The final parameter in the serial interface menu is the linefeed mode. This pa-
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
LF= On
Adjust the setting using "UP" or "DOWN" and pressing "SET".
LF=OFF
7.13

Calibration Parameters

The operator of the instrument controller has access to a number of the calibra-
tion constants namely R0, ALPHA, and DELTA. These values are set at the
factory and must not be altered. The correct values are important to the accu-
racy and proper and safe operation of the instrument. Access to these parame-
ters is available to the user so that in the event that the controller memory fails
the user may restore these values to the factory settings. The user should have a
list of these constants and their settings with the instrument manual.
Caution: DO NOT change the values of the instrument calibration con-
stants from the factory set values. The correct setting of these parameters
is important to the safety and proper operation of the instrument.
The calibration parameters menu is indicated by,
Press "SET" five times to enter the menu. The calibration parameters menu
contains the parameters, R0, ALPHA, and DELTA, which characterize the re-
sistance-temperature relationship of the platinum control sensor. These parame-
ters may be adjusted to improve the accuracy of the calibrator.
The name of the parameter flashes on the display and then the value is dis-
played. The value of the parameter may be changed using the "UP" and
"DOWN" buttons. After the desired value is reached press "SET" to set the pa-
New duplex mode setting
LF
Flashes LF and then displays the setting
Current linefeed setting
New linefeed setting
CAL
Calibration parameters menu
7 Controller Operation
Calibration Parameters
29