Endress+Hauser Cerabar S PMC71 Technical Information

Pressure transmitter with ceramic and metal sensors overload-resistant and function-monitored; communication via hart profibus pa or foundation fieldbus
Hide thumbs Also See for Cerabar S PMC71:
Table of Contents

Advertisement

Technical Information
Cerabar S PMC71, PMP71/72/75
Pressure transmitter
with ceramic and metal sensors
Overload-resistant and function-monitored; Communication via
HART, PROFIBUS PA or FOUNDATION Fieldbus
Application
The Cerabar S pressure transmitter is used for the
following measuring tasks:
• Absolute pressure and gauge pressure in gases, steams
or liquids in all areas of process engineering and
process measurement technology
• Level, volume or mass measurement in liquids
• High process temperature
– without diaphragm seals up to 150°C (302°F)
– with typical diaphragm seals up to 350°C (662°F)
• High pressure up to 700 bar
• International usage thanks to a wide range of approvals
TI383P/00/en
Your benefits
• Very good reproducibility and long-term stability
• High reference accuracy: up to ±0.075%,
as PLATINUM version: ±0.05%
• Turn down 100:1, higher on request
• Meets PED (Pressure Equipment Directive)
• Used for process pressure monitoring up to SIL 2,
certified according to IEC 61508 by TÜV SÜD
®
• HistoROM
/M-DAT memory module
• Function-monitored from the measuring cell to the
electronics
• Continuous modularity for differential pressure and
pressure (Deltabar S – Cerabar S), e.g.
– replaceable display
– universal electronic
• Quick commissioning thanks to quick setup menu
• Easy and safe menu-guided operation on-site,
via 4...20 mA with HART, via PROFIBUS PA or
via FOUNDATION Fieldbus
• Extensive diagnostic functions
• Device versions in conformity with ASME-BPE

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Cerabar S PMC71

  • Page 1 Technical Information Cerabar S PMC71, PMP71/72/75 Pressure transmitter with ceramic and metal sensors Overload-resistant and function-monitored; Communication via HART, PROFIBUS PA or FOUNDATION Fieldbus Application Your benefits The Cerabar S pressure transmitter is used for the • Very good reproducibility and long-term stability •...
  • Page 2: Table Of Contents

    Material ......... . . 61 Endress+Hauser...
  • Page 3 Installation/Control Drawings ......83 Overspill protection ........83 Endress+Hauser...
  • Page 4: Function And System Design

    OPL = Over pressure limit; dependent on the lowest-rated element, with regard to pressure, of the selected components High temperature version "T" for feature 100 "Additional option 1" or for feature 110 "Additional option 2" lower temperature on request Endress+Hauser...
  • Page 5: Overview Of Diaphragm Seal For Pmp75

    DN 50/DN 80 + PN 10 – PN 40 extended diaphragm seal DIN 2527 50/100/200 mm ext. diaphragm seal (MDM) diaphr. seal ANSI flange ANSI B 16.5 2"/3"/4" + 2"/4"/6"/ Up to 300 lbs P01-PMP75xxx-03-xx-xx-xx-002 8" ext. diaphr. seal Endress+Hauser...
  • Page 6 EN837 – G 1/2 B – 400 bar separator (MDM) P01-PMP75xxx-03-xx-xx-xx-004 ANSI – 1/2 MNPT – 160 bar – 400 bar P01-PMP75xxx-03-xx-xx-xx-008 NPT Off line thread ANSI – 1/2 NPT 250 bar – 1 NPT P01-PMP75xxx-03-xx-xx-xx-009 Endress+Hauser...
  • Page 7: Measuring Principle

    Advantages: • Can be used with process pressures from 400 mbar to 400 bar • High long-term stability • Guaranteed overload resistance up to 4 times the nominal pressure • Second process barrier (Secondary Containment) for enhanced integrity Endress+Hauser...
  • Page 8: Level Measurement (Level, Volume And Mass)

    • 4...20 mA with HART communication protocol • PROFIBUS PA – The Endress+Hauser devices meet the requirements as per the FISCO model. – Due to the low current consumption of 11 mA ± 1 mA – up to 9 Cerabar S for EEx ia, CSA IS and FM IS applications –...
  • Page 9: Human Interface

    • Comprehensive diagnostic functions (fault and warning message) Parameter Measured value display Identification Function name Value number Header line Main line Unit Information line Symbol Bargraph P01-xMD7xxxx-07-xx-xx-xx-001 Endress+Hauser...
  • Page 10: Operating Elements

    ® measured-value-relevant parameters Slot for optional HistoROM /M-DAT DIP-switch for locking/unlocking DIP-switch for damping on/off DIP-switch for locking/unlocking measured- measured-value-relevant parameters Green LED to indicate value being accepted value-relevant parameters DIP-switch for simulation mode on/off DIP-switch for damping on/off Endress+Hauser...
  • Page 11: Historom /M-Dat (Optional)

    /M-DAT can be ordered via feature 100 "Additional options 1" or feature 110 "Additional options 2" or as spare parts. → See also page 70 ff. A CD with the Endress+Hauser ToF Tool operating program is also included in the scope of delivery.
  • Page 12: Tof Tool - Hart, Profibus Pa, Foundation Fieldbus

    Remote operation – An FF configuration program is required to integrate a device with "FOUNDATION Fieldbus signal" into an FF FOUNDATION Fieldbus network or to set the FF-specific parameters. Please contact your local Endress+Hauser Sales Center for more information. Service interface FXA193...
  • Page 13: Input

    For the appropriate standards and further information, see page 33, "Pressure specification". Versions in the order code → See also page 71 ff, feature 40 "Sensor range; Sensor overload limit (= OPL)" minimum span that can be calibrated, Turn down > 100:1 on request Endress+Hauser...
  • Page 14 PMP75. → See also page 63, section "Diaphragm seal filling oils". Versions in the order code → See also page 70 ff, feature 40 "Sensor range; Sensor Overload limit (= OPL)" minimum span that can be calibrated, Turn down > 100:1 on request PMP71 only, PMP75 on request Endress+Hauser...
  • Page 15: Explanation Of Terms

    = 0.6 bar Set span This span is based on the zero point. Zero based span Nominal value i Upper range limit (URL) Nominal measuring range Sensor measuring range Lower range limit Upper range limit Lower range value Upper range value Endress+Hauser...
  • Page 16: Output

    Supply voltage 11 (12)...45 V DC for PMC71, EEx d[ia], NEPSI Ex d[ia] and TIIS Ex d[ia] Maximum load resistance Lmax Supply voltage Note! When operating via a handheld terminal or via PC with an operating program, a minimum communication resistance of 250 Ω must exist within the loop. Endress+Hauser...
  • Page 17: Resolution

    • Via on-site display, handheld terminal or PC with operating program, continuous from 0...999 s • Additionally for HART and PROFIBUS PA: via DIP-switch on the electronic insert, switch position "on" = set value and "off" • Factory setting: 2 s Endress+Hauser...
  • Page 18: Power Supply

    The two-wire cable must be connected to the "FF+" and "FF–" terminals. Devices with Harting plug Han7D Cerabar S + – – Han7D – P01-PMx7xxxx-04-xx-xx-xx-001 Left: electrical connection for devices with Harting plug Han7D Right: view of the plug at the device Endress+Hauser...
  • Page 19 P01-PMx7xxxx-04-xx-xx-xx-000 Left: electrical connection for devices with M12 plug Right: view of the plug at the device Endress+Hauser offers for devices with M12 plug the following accessories: Plug-in jack M 12x1, straight • Material: Body PA; coupling nut CuZn, nickel-plated •...
  • Page 20: Supply Voltage

    → See also page 70 ff, feature 30 "Housing, Cable entry, Protection". Cable entry Cable specification • Endress+Hauser recommends using shielded, twisted-pair two-wire cables. • Terminals for wire cross-sections 0.5...2.5 mm • Cable external diameter: 5...9 mm Without influence on 4...20 mA signal up to ± 5% residual ripple within the permitted voltage range [according...
  • Page 21: Performance Characteristics - General

    Device rotated 180°, process connection pointing upwards. This value is doubled for devices with inert oil. Position-dependent zero shift can be corrected. → See also page 27, section "General installation instructions" and page 67 ff section "Installation instructions, Diaphragm seal systems". Endress+Hauser...
  • Page 22: Performance Characteristics - Ceramic Diaphragm

    1 bar, 2 bar, 4 bar, 10 bar, 40 bar measuring cell: • ±0.15 % of URL PMC71 High temperature version all measuring cells: • ±0.46% of URL Warm-up period – • 4...20 mA HART : < 10 s PMC71 • PROFIBUS PA: 6 s • FOUNDATION Fieldbus: 50 s Endress+Hauser...
  • Page 23: Dead Time, Time Constant (T63) - Pmc71

    • 100 mbar, 250 mbar, 400 mbar measuring cell: ±(0.088 x TD + 0.088)% of the set span • 1 bar, 2 bar, 4 bar, 10 bar, 40 bar measuring cell: ±(0.088 x TD + 0.04)% of the set span –20...–10°C, +60...+150°C (–4...+14°F, +140...+302°F): • all measuring cells: ±1.25% of URL x TD Endress+Hauser...
  • Page 24: Performance Characteristics - Metallic Diaphragm

    • TD 1:1 to TD 2:1: ±0.2% of the set span • TD > 2:1: ±0.1% of the set span x TD Platinum version, 2 bar, 4bar, 10 bar, 40 bar measuring cell: • TD 1:1: ±0.05% of the set span Endress+Hauser...
  • Page 25: Total Performance - Pmp71

    400 bar measuring cell: • ±0.3% of URL 700 bar measuring cell: • ±0.4% of URL Warm-up period – • 4...20 mA HART : < 10 s PMP71, PMP75 • PROFIBUS PA: 6 s • FOUNDATION Fieldbus: 50 s Endress+Hauser...
  • Page 26: Dead Time, Time Constant (T63) - Pmp71, Pmp75

    • 100 bar measuring cell: ±(0.4 x TD + 0.03)% of the set span • 400 bar measuring cell: ±(0.7 x TD + 0.04)% of the set span • 700 bar measuring cell: ±(0.75 x TD + 0.06)% of the set span Endress+Hauser...
  • Page 27: Operating Conditions (Installation)

    • The housing of the Cerabar S can be rotated up to 380°. → See also page 29, section "Turn the housing". • Endress+Hauser offers a mounting bracket for installing on pipes or walls. → See also page 28, section "Wall and pipe-mounting".
  • Page 28: Mounting With Temperature Isolator

    The position-dependent zero shift can be corrected. P01-PMx7xxxx-11-xx-xx-xx-005 PMP75 with temperature isolator Endress+Hauser offers a mounting bracket for installing the device on pipes or walls. → See also page 70 ff, Wall and pipe-mounting feature 110, "Additional options 2".
  • Page 29: Turn The Housing

    (1.5 x PN) or filling fluid (160 bar) → See page 70 ff "Ordering information", feature 40 "Sensor range; sensor overload limit (= OPL)" PMC71 with PVDF thread or flange p = 15 bar (225 psi) Endress+Hauser...
  • Page 30: Ultra Pure Gas Applications

    Cerabar S Ultra pure gas applications Endress+Hauser also offers the degreased device for special applications, such as ultra pure gas. No special restrictions regarding the process conditions apply to this device. → See also page 72, "Ordering information PMC71", feature 80 "Seal" or page 76, "Ordering information PMP71", feature 90 "Fill fluid".
  • Page 31: Vibration Resistance

    For applications with high temperatures, either a PMP75 with a temperature isolator or a PMP75 with a capillary can be used. If vibrations also occur in the application, Endress+Hauser recommends using a PMP75 with a capillary. If a PMP75 with a temperature isolator is used, it must be mounted with a mounting bracket. (→ see also page 28).
  • Page 32: Operating Conditions (Process)

    These seals are used for devices with 3A-approved process connections. For devices with NBR or HNBR seals, the values for "Toatal Performance" (→ see page 22) and "Thermal change" (→ see page 23) must be multiplied by the factor 3. Endress+Hauser...
  • Page 33: Pressure Specifications

    With regard to its stability property, the material 1.4435 is identical to 1.4404 which is grouped under 13EO in EN 1092-1 Tab. 18. the chemical composition of the two materials can be identical. The equation does not apply for PMP71 and PMP75 with a 100 bar measuring cell. Endress+Hauser...
  • Page 34: Mechanical Construction

    → See the process connection in question for installation height. Housing weight see page 61. Housing dimensions T17 P01-PMx7xxxx-06-00-xx-xx-001 Front view, left-hand side view, top view → See the process connection in question for installation height. Housing weight see page 61. Endress+Hauser...
  • Page 35: Process Connections Pmc7

    Thread ANSI 1/2 MNPT hole 3 mm; Material version RG: PVDF(max.: 15 bar/225 psi, max.: –10...+60°C/+14...+140°F), mount with mounting bracket only (→ see also page 28) Thread ANSI 1/2 FNPT; Material version RH: AISI 316L/1.4435, version RJ: Alloy C276/2.4819, version RK: Monel Endress+Hauser...
  • Page 36 251 mm PMC 71 High temperature version with 300 mm 316 mm EEx d[ia], CSA XP or FM XP High temperature version, see also page 73, feature 100 "Additional options 1" and feature 110 "Additional options 2", versions "T" Endress+Hauser...
  • Page 37 Material version 1R: AISI 316L/1.4435, version 1S: Alloy C276/2.4819 Installation height H for devices with thread connection and flush-mounted diaphragm Description Housing T14 Housing T17 PMC71 215 mm 231 mm PMC71 with EEx d[ia], CSA XP or FM XP 280 mm 296 mm Endress+Hauser...
  • Page 38 B2 (D) PVDF DN 80 PN 10/16 B1 (D) 21.4 AISI 316L DN 80 PN 10-40 B1 (D) Designation in brackets as per DIN 2527 Housing weight see page 61 AISI 316L/1.4435 Max.: 15 bar (225 psi), max.: –10...+60°C (+14...+140°F) Endress+Hauser...
  • Page 39 Combination of AISI 316 for required pressure resistance and AISI 316L for required chemical resistance (dual rated) ECTFE coating on AISI 316L/1.4435 When operating in hazardous area, avoid electrostatic charge of the plastic surface. max.: 15 bar (225 psi), max.: –10...+60°C (+14...+140°F) Endress+Hauser...
  • Page 40 100 A 10 K Housing weight see page 61 Installation height H for devices with flange Description T14 housing T17 housing PMC71 215 mm 231 mm PMC71 with EEx d[ia], CSA XP or FM XP 280 mm 296 mm Endress+Hauser...
  • Page 41 Version TR: Varivent Type N for pipes 40 – 162, PN 40, 3A with HNBR or EPDM seal Installation height H for devices with hygienic connection and flush-mounted diaphragm Description T14 housing T17 housing PMC71 215 mm 231 mm PMC71 with EEx d[ia], CSA XP or FM XP 280 mm 296 mm Endress+Hauser...
  • Page 42: Process Connections Pmp71

    Thread ANSI 1/2 MNPT 1/4 FNPT; Material version RA: AISI 316L/1.4435, version RB: Alloy C276/2.4819 Thread ANSI 1/2 MNPT hole 11.4; Material version RD: AISI 316L/1.4435, version RE: Alloy C276/2.4819 Thread ANSI 1/2 FNPT; Material version RH: AISI 316L/1.4435, version RJ: Alloy C276/2.4819 Endress+Hauser...
  • Page 43 Process connections PMP71 thread DIN 13 M 20x1.5 hole 11.4 mm Material version GP: AISI 316L/1.4435, version GQ: Alloy C276/2.4819 → Installation height H see table, below. Installation height H for devices with thread connection and internal flush-mounted diaphragm T14 housing T17 housing Height H 165 mm 181 mm Endress+Hauser...
  • Page 44 Thread ANSI 1 MNPT; Material version 2A: AISI 316L/1.4435, version 2B: Alloy C276/2.4819 Thread ANSI 1 1/2 MNPT; Material version 2D: AISI 316L/1.4435, version 2E: Alloy C276/2.4819 Thread ANSI 2 MNPT Material version 2G: AISI 316L/1.4435, version 2H: Alloy C276/2.4819 Endress+Hauser...
  • Page 45 179 mm G 2 A 162 mm 178 mm 1 MNPT 162 mm 178 mm 1 1/2 MNPT 165 mm 181 mm 2 MNPT 159 mm 175 mm M 20x1.5 163 mm 179 mm M 44x1.25 170 mm 186 mm Endress+Hauser...
  • Page 46 PN 10-40 B1 (D) DN 32 PN 10-40 B1 (D) DN 40 PN 10-40 B1 (D) DN 50 PN 10-40 B1 (D) DN 80 PN 10-40 B1 (D) Designation as per DIN 2527 in brackets Housing weight see page 61 Endress+Hauser...
  • Page 47 50 A 10 K AISI 316L 80 A 10 K AISI 316L 100 A 10 K Housing weight see page 61 Combination of AISI 316 for required pressure resistance and AISI 316L for required chemical resistance (dual rated) AISI 316L/1.4435 Endress+Hauser...
  • Page 48 Version UR: oval flange adapter 1/4-18 NPT, mounting: 7/16-20 UNF Prepared for diaphragm seal mount ø2.5 ø7.95 P01-PMP75xxx-06-09-xx-xx-xx-005 Version U1: prepared for diaphragm seal mount Hole for filling fluid Bearing Threaded pin with an internal hexagon 4 mm T14 housing T17 housing Height h 190 mm 204 mm Endress+Hauser...
  • Page 49: Pmp75 Basic Unit

    • The weights of the diaphragm seals are given in the tables. See page 61 for the weight of the housing. • The following drawings are drawings that illustrate how the system works in principle. In other words, the dimensions of a diaphragm seal supplied can deviate from the dimensions given in this document. Endress+Hauser...
  • Page 50 +24.42 +4.21 Alloy C276 – – – AISI 316L 1 1/2 MNPT 400 – +8.14 +12.39 +2.59 Alloy C276 – – – AISI 316L 2 MNPT – +5.4 +8.18 +2.59 Alloy C276 – – – AISI 316L/1.4435; Alloy C276/2.4819 Endress+Hauser...
  • Page 51 +0.35 +0.18 ≤ 0.4 µm (15.75 µin; 180 grit), electropolished; Diaphragm seal versions in conformity with ASME-BPE for use in biochemical processes, wetted surfaces R to be ordered using feature 60 "Additional option", version "P" in the order code Endress+Hauser...
  • Page 52 +3.11 +0.89 1.4435 Diaphragm seal versions in conformity with ASME-BPE for use in biochemical processes, wetted surfaces ≤ 0.4 µm (15.75 µin; 180 grit), electropolished; to be ordered using feature 60 "Additional option", version "P" in the order code Endress+Hauser...
  • Page 53 [mm] [mm] [mm] [mbar/10 K] [kg] PN 25 Rd 40 – 1/6 42.5 +15.66 +24.22 +4.21 0.25 1 1/2 PN 25 Rd 60 – 1/6 +8.18 +12.39 +2.59 0.65 PN 25 Rd 70 – 1/6 +5.4 +8.18 +1.76 1.05 Endress+Hauser...
  • Page 54 [bar] [mm] [mm] [mm] [mm] [mm] [mbar/10 K] [kg] PN 40 54.1 1 1/2" – 1/8" 46.8 +15.66 +24.42 +4.21 1 1/2 PN 40 2" – 1/8" +8.14 +12.39 +2.59 PN 40 2 1/2" – 1/8" +5.4 +8.18 +1.76 Endress+Hauser...
  • Page 55 [mm] [mm] [mbar/10 K] [kg] DN 50 PN 25 Rd 78 x 1/6" +2.21 +3.02 +0.88 DN 65 PN 25 Rd 95 x 1/6" +1.6 +2.1 +0.6 DN 80 PN 25 Rd 110 x 1/4" 40 +0.66 +0.81 +0.4 Endress+Hauser...
  • Page 56 Extension Diaphragm seal diameter pressure length diameter weight [mm] [mm] [kg] DN 50 PN 10-40 50 / 100 / 200 48.3 3.2 / 3.8 / 4.4 DN 80 PN 10-40 50 /100 / 200 6.2 / 6.7 / 7.8 Endress+Hauser...
  • Page 57 3.62 1.12 6.75 2.32 +2.21 +3.02 +1.15 15.8 316L 50.8 91.9 28.4 171.5 AISI 316/ 0.94 0.75 3.50 +0.19 +0.25 +0.11 316L 190.5 23.9 19.1 152.4 AISI 316/ 8.25 1.12 0.75 +0.19 +0.25 +0.11 316L 209.5 28.4 19.1 152.4 Endress+Hauser...
  • Page 58 – 8.6 – 4 (101.6) – 9.9 – 6 (152.4) – 11.2 – 8 (203.2) – 12.4 – 2 (50.8) 3.7 (94) – 13.1 – 4 (101.6) – 14.4 – 6 (152.4) – 15.7 – 8 (203.2) – 16.9 Endress+Hauser...
  • Page 59 Separator ISO 228 G 1/2 A and ANSI 1/2 MNPT ø72.6 ø72.6 ø72.6 SW 30 SW 30 AF 30 AF 30 ø18.4 ø11.4 ø4 ø6 1/2 - 14 NPT G 1/2A P01-PMP75xxx-06-09-xx-xx-004 Process connection PMP75, versions "UA" and "UB", welded, material AISI 316L Endress+Hauser...
  • Page 60 1/2 NPT und 1 NPT Off line thread ø118 ø118 1/2 NPT SW 27 27 AF 1 NPT ø30 SW 36 36 AF ø52 P01-PMP75xxx-06-09-xx-xx-008 Process connection PMP75, versions "UG" and "UL", screwed, material AISI 316L , seal Viton Endress+Hauser...
  • Page 61: Weight

    • Mounting accessories: Mounting kit with screws AISI 304 (1.4301) • Capillary: AISI 316 Ti (1.4571) • Protective hose for capillary: AISI 304 (1.4301) • External earth terminal: AISI 304 (1.4301) → For process connections, process diaphragms, seals and filling oils see ordering information, page 70 ff. Endress+Hauser...
  • Page 62: Planning Instructions, Diaphragm Seal Systems

    The process pressure acts via the diaphragm seal membrane on the liquid-filled system, which transfers the process pressure via the capillary tube onto the sensor of the pressure transmitter. Endress+Hauser delivers all diaphragm seal systems as welded versions. The system is hermetically sealed, which ensures the highest reliability.
  • Page 63: Diaphragm Seal Filling Oils

    The volume change is the volume that has to be shifted to pass through the complete measuring range. Pressure transmitters from Endress+Hauser are optimised with regard to minimum volume change. Diaphragm seal filling oils...
  • Page 64 • The influence of temperature on the zero point can be corrected with position calibration. • The temperature influence can be minimised by using a filling oil with a smaller coefficient of thermal expansion, shorter capillary, diaphragm seal with larger diaphragm diameter or by using a smaller capillary internal diameter. Endress+Hauser...
  • Page 65 DIN 11851 DN 50 PN 25 socket, AISI 316L/1.4435 DIN 11851 DN 65 PN 25 socket, AISI 316L/1.4435 Varivent Type N for tubes DN 40 – DN 162, PN 40, AISI 316L/1.4435 DRD 65 mm, PN 25, AISI 316L/1.4435 Endress+Hauser...
  • Page 66 Ambient dependent on the capillary length for PMP75, PN ≤ 40 bar Diagram T PN > 40 bar T Ambient [mbar/10 K] Capillary length P01-PMP75xxx-05-xx-xx-xx-009 Ambient dependent on the capillary length for PMP75, PN > 40 bar Diagram T Endress+Hauser...
  • Page 67: Ambient Temperature Range

    In order to obtain more precise measurement results and to avoid a defect in the device, mount the capillaries as follows: • vibration-free (in order to avoid additional pressure fluctuations) • not in the vicinity of heating or cooling lines • insulate at colder or warmer ambient temperatures • with a bending radius of ≥ 100 mm. Endress+Hauser...
  • Page 68 Cerabar S Vacuum applications For applications under vacuum, Endress+Hauser recommends mounting the pressure transmitter below the diaphragm seal. A vacuum load of the diaphragm seal caused by the presence of fill fluid in the capillary prevents is hereby prevented. When the pressure transmitter is mounted above the diaphragm seal, the maximum height difference H1 in accordance with the following illustration on the left must not be exceeded.
  • Page 69: Certificates And Approvals

    Cerabar S Certificates and approvals CE mark The device meets the legal requirements of the relevant EC directives. Endress+Hauser confirms that the device has been successfully tested by applying the CE mark. Ex approvals • ATEX • FM • CSA •...
  • Page 70: Ordering Information

    AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, M 12x1 PA plug AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, 7/8" FF plug PMC71 order code → For continuation of ordering information for PMC71, see the following page. Endress+Hauser...
  • Page 71 Thread DIN 13 M 20x1.5 EN 837 hole 3 mm, AISI 316L Thread DIN 13 M 20x1.5 EN 837 hole 3 mm, Alloy C For continuation "Process connection, Material" see next page. PMC71 order code → For continuation of ordering information for PMC71, see the following page. Endress+Hauser...
  • Page 72 Varivent type N for tubes DN 40 – DN 162, PN 40, AISI 316L, 3A with HNBR/EPDM seal Seal: FKM Viton EPDM Kalrez Chemraz NBR/3A: HNBR (FDA) FKM Viton, degreased FKM Viton, cleaned for oxygen service PMC71 order code → For continuation of ordering information for PMC71, see the following page. Endress+Hauser...
  • Page 73 Routine test with certificate, inspection certificate as per EN 10204 3.1 Overpressure test with certificate, inspection certificate as per EN 10204 3.1 Helium leak test EN 1518 with test certificate, inspection certificate as per EN 10204 3.1 PMC71 complete order code Endress+Hauser...
  • Page 74: Pmp71

    AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, M 12x1 PA plug AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, 7/8" FF plug PMP71 order code → For continuation of ordering information for PMP71, see the following page. Endress+Hauser...
  • Page 75 Thread JIS B0202 G 1/2 (male), AISI 316L Thread JIS B0203 R 1/2 (male), AISI 316L For continuation "Process connection; Material", see next page. PMP71 order code → For continuation of ordering information for PMP71, see the following page. Endress+Hauser...
  • Page 76 Prepared for diaphragm seal mount, AISI 316L Fill fluid: Silicone oil fill Inert oil fill Inert oil fill, degreased Inert oil fill, cleaned for oxygen services PMP71 order code → For continuation of ordering information for PMP71, see the following page. Endress+Hauser...
  • Page 77 Routine test with certificate, inspection certificate as per EN 10204 3.1 Overpressure test with certificate, inspection certificate as per EN 10204 Helium leak test EN 1528 with test certificate, inspection certificate as per EN 10204 3.1 PMP71 complete order code Endress+Hauser...
  • Page 78: Pmp75

    AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, M 12x1 PA plug AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, 7/8" FF plug PMP75 order code → For continuation of ordering information for PMP75, see the following page. Endress+Hauser...
  • Page 79 Tri-Clamp, ISO 2852 DN 38 (1 1/2"), DIN 32676 DN 40, AISI 316L Tri-Clamp, ISO 2852 DN 40 – DN 51 (2")/DN 50, AISI 316L Tri-Clamp, ISO 2852 DN 70 – DN 76.1 (3"), AISI 316L PMP75 order code → For continuation of ordering information for PMP75, see the following page. Endress+Hauser...
  • Page 80 3" 150 lbs RF, AISI 316/316L (CRN) 3" 300 lbs RF, AISI 316/316L (CRN) 4" 150 lbs RF, AISI 316/316L (CRN) 4" 300 lbs RF, AISI 316/316L (CRN) PMP75 order code → For continuation of ordering information for PMP75, see the following page. Endress+Hauser...
  • Page 81 Mounting bracket for wall/pipe, AISI 304 Test report acc. to EN 10204 2.2 Routine test with certificate, inspection certificate as per EN 10204 3.1 Overpressure test with certificate, inspection certificate as per EN 10204 3.1 PMP75 complete order code Endress+Hauser...
  • Page 82: Documentation

    – 4...20 mA HART, – XA276P PMP72, PMP75 PROFIBUS PA, FOUNDATION Fieldbus ATEX II 1/2 G EEx ia IIC T6 PMP71, PMP75 – 4...20 mA HART, – XA252P + ATEX II 2 G EEx d IIC T6 PROFIBUS PA, FOUNDATION Fieldbus Endress+Hauser...
  • Page 83: Installation/Control Drawings

    FOUNDATION Fieldbus FM/CSA IS + XP Class I, PMP71, PMP72, – 4...20 mA HART – ZD154P + ZD187P Division 1, Groups A – D PMP75 – PROFIBUS PA, – ZD190P + ZD191P FOUNDATION Fieldbus Overspill protection • WHG: ZE260P/00/de Endress+Hauser...
  • Page 84 Cerabar S International Head Quarter Endress+Hauser GmbH+Co. KG Instruments International Colmarer Str. 6 79576 Weil am Rhein Deutschland Tel. +49 76 21 9 75 02 Fax +49 76 21 9 75 34 5 www.endress.com info@ii.endress.com TI383P/00/en/03.06 CCS/FM+SGML 6.0...

This manual is also suitable for:

Cerabar s pmp75Cerabar s pmp72Cerabar s pmp71

Table of Contents