Operating Instructions – Description of Device Functions
Cerabar S
PMC71, PMP71/75
Deltabar S
FMD76/77/78, PMD70/75
Deltapilot S FMB70
Process pressure / Differential pressure, Flow / Hydrostatic
BA00303P/00/EN/13.11
71139808
valid from Software version:
03.00.zz

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Cerabar S PMC71

  • Page 1 Operating Instructions – Description of Device Functions Cerabar S PMC71, PMP71/75 Deltabar S FMD76/77/78, PMD70/75 Deltapilot S FMB70 Process pressure / Differential pressure, Flow / Hydrostatic BA00303P/00/EN/13.11 71139808 valid from Software version: 03.00.zz...
  • Page 2 Brief Operating Instructions – Installation KA01024P – Wiring – Local operation – Commissioning – Description of Quick Setup menus Fold-out brochure KA00252P – Wiring – Operation without display – Description of Quick Setup menus ® – HistoROM /M-DAT operation Endress+Hauser...
  • Page 3 Brief Operating Instructions – Installation KA01026P – Wiring – Local operation – Commissioning – Description of Quick Setup menus Fold-out brochure KA00252P – Wiring – Operation without display – Description of Quick Setup menus ® – HistoROM /M-DAT operation Endress+Hauser...
  • Page 4 Overview of documentation Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus Endress+Hauser...
  • Page 5: Table Of Contents

    Overview of level measurement ....105 "Level easy pressure" level selection ... 106 "Level easy height" level selection ... . 110 Endress+Hauser...
  • Page 6: Notes On Use

    PID or Discret Ouput Block, see Operating Instructions BA00013S "FOUNDATION Fieldbus Overview" or the FOUNDATION Fieldbus Specification. Sections 8 to 10 describe how to operate the unit via the local operation and the Endress+Hauser operating program FieldCare.
  • Page 7: Parameter Description Of Local Operation And Fieldcare

    TEMP. ENG. UNIT – "Pressure" measuring mode TEMP. ENG. UNIT – "Level" measuring mode TEMP. ENG. UNIT – "Flow" measuring mode CALIB. OFFSET SET. L. FL. CUT-OFF FACT. U.U. TOTAL.1 FACT. U.U. TOTAL.2 RESET TOTALIZER 1 Pmin ALARM WINDOW Pmax ALARM WINDOW Endress+Hauser...
  • Page 8 PCB COUNT: T < Tmin PCB MIN. TEMP. ACK. ALARM MEASURING TABLE (display) EDITOR TABLE, LINE-NUMB (enter values) EDITOR TABLE, X-VAL. (enter values) EDITOR TABLE, Y-VAL. (enter values) 187, 188 POS. INPUT VALUE 137, 139 or 143 LAST DIAG. CODE Pmax PROC. CONN. Endress+Hauser...
  • Page 9 FULL PRESSURE – "Linear" level mode FULL PRESSURE – "Height linearized" level mode LEVEL MAX. TANK CONTENT MAX. SIM. LEVEL SIM. TANK CONT. MEASURING TABLE (selection) LEVEL MODE LEVEL MIN. TANK CONTENT MIN. HYDR. PRESS MAX. EDITOR TABLE (continue entries) Endress+Hauser...
  • Page 10 ZERO POSITION – "Linear" level mode ZERO POSITION – "Height linearized" level mode TANK DESCRIPTION HistoROM AVAIL. HistoROM CONTROL TANK VOLUME TANK HEIGHT AI 3 OUT Value AI 2 OUT Value AI 1 OUT Value DEVICE ADDRESS DD REVISION DEVICE REVISION DEVICE ID Endress+Hauser...
  • Page 11: Graphic Representation Of Function Groups 11

    → 201 → PROCESS VALUES (flow) * → 202 → FUNCTION BLOCKS → 203 → PEAK HOLD INDICATOR → 203 → OPERATION → 205 → DIAGNOSTICS → SIMULATION MODE → 206 → MESSAGES → 207 → USER LIMITS → 208 Endress+Hauser...
  • Page 12: Representation In Fieldcare

    → 201 → PROCESS VALUES (flow) * → 202 → FUNCTION BLOCKS → 203 → PEAK HOLD INDICATOR → 203 → OPERATION → 205 → DIAGNOSTICS → SIMULATION MODE → 206 → MESSAGES → 207 → USER LIMITS → 208 Endress+Hauser...
  • Page 13: Pressure Measurement (Ff Configuration Program)

    Analog Input Block. Result: The device is ready for pressure measurement. Note! • You can select another pressure unit by means of the Calibration Units/CAL_UNIT parameter (→ ä 54). You can also specify a customer-specific unit by means of this parameter. Endress+Hauser...
  • Page 14: Level Measurement (Ff Configuration Program)

    % must not be in direct proportion to the measured pressure. A linearization table has to be entered for the 2nd measured variable. The 2nd measured variable is assigned to the 1st measured variable by means of this table. Endress+Hauser...
  • Page 15: Level Easy Pressure" Level Selection

    See Operating Instructions 300 mbar BA00301P. Open the Pressure Transducer Block and set the block mode to OOS. ➀ 0 mbar P01-PMP75xxx-19-xx-xx-xx-008 Fig. 1: Calibration with reference pressure – wet calibration See Table, Step 8. See Table, Step 10. Endress+Hauser...
  • Page 16 Set the Pressure Transducer Block to the "Auto" block mode. Where necessary, configure the Channel/CHANNEL (→ ä 96), Linearization Type/L_TYPE (→ ä 97), Transducer Scale/XD_SCALE (→ ä 95) and Output Scale/OUT_SCALE (→ ä 96) parameters by means of the Analog Input Block. Endress+Hauser...
  • Page 17 – Pos.Zero Adjust/ZERO_POSITION_ADJUST (→ ä 59) or – Pos. Input Value/POSITION_INPUT_VALUE (→ ä 59) or – Calib. Offset/CALIBRATION_OFFSET (→ ä 59). P01-PMC71xxx-19-xx-xx-xx-008 Fig. 3: Calibration without reference pressure – dry calibration See Table, Step 10. See Table, Step 9. Endress+Hauser...
  • Page 18 Set the Pressure Transducer Block to the "Auto" block mode. Where necessary, configure the Channel/CHANNEL (→ ä 96), Linearization Type/L_TYPE (→ ä 97), Transducer Scale/XD_SCALE (→ ä 95) and Scale Out/ SCALE OUT (→ ä 96) parameters by means of the Analog Input Block. Endress+Hauser...
  • Page 19: Level Easy Height" Level Selection

    Open the Pressure Transducer Block and set the block mode to OOS. ➁ 0.5 m ➀ r = 1 kg P01-PMC71xxx-19-xx-xx-xx-009 Fig. 5: Calibration with reference pressure – wet calibration See Table, Steps 10 and 11. See Table, Step 12. See Table, Step 13. Endress+Hauser...
  • Page 20 Set the Pressure Transducer Block to the "Auto" block mode. Where necessary, configure the Channel/CHANNEL (→ ä 96),Linearization Type/L_TYPE (→ ä 97), Transducer Scale/XD_SCALE (→ ä 95) and Output Scale/OUT_SCALE (→ ä 96) parameters by means of the Analog Input Block. Endress+Hauser...
  • Page 21 – Pos. Input Value/POSITION_INPUT_VALUE (→ ä 59) or – Calib. Offset/CALIBRATION_OFFSET (→ ä 59). P01-PMC71xxx-19-xx-xx-xx-009 Fig. 7: Calibration without reference pressure – dry calibration See Table, Step 10. See Table, Steps 12 and 13. See Table, Steps 14 and 15. Endress+Hauser...
  • Page 22 See Table, Step 13. See Table, Step 14 By means of the Full Height Level Easy/ See Table, Step 15. FULL_HEIGHT_EASY parameter, enter a volume, here 4.5 m (15 ft) for example. Set the Pressure Transducer Block to the "Auto" block mode. Endress+Hauser...
  • Page 23: Level Standard" Level Selection, "Linear" Level Mode

    By means of the Calibration mode/CALIBRATION_ MODE parameter, select the "Wet" option. See also Point 3 in the following note. Fill the container up to the lower level point. The related pressure value can be viewed via the Pressure/PRESSURE parameter. Endress+Hauser...
  • Page 24 (→ ä 96), Linearization Type/L_TYPE (→ ä 97), Transducer Scale/XD_SCALE (→ ä 95) and Output [mbar] Scale/OUT_SCALE (→ ä 96) parameters by means of the Analog Input Block. P01-xxxxxxxx-05-xx-xx-xx-011 1 See Table, Step 11. 2 See Table, Step 13. Endress+Hauser...
  • Page 25 "Linear" option. By means of the Lin. Measurand/LINEAR_ MEASURAND parameter, select the "Volume" option. Select a volume unit via the Unit Volume/ VOLUME_UNIT parameter, here m for example. By means of the Calibration Mode/ CALIBRATION_MODE parameter, select the "Dry" option. Endress+Hauser...
  • Page 26 For this level mode, the measured variables %, level, volume and mass are available. → ä 61 ff. You can also specify customer-specific units. See parameter description for Calibration Units/ CAL_UNIT (→ ä 54), Height Unit/HEIGHT_UNIT (→ ä 62), Unit Volume/ VOLUME_UNIT (→ ä 64) and Mass Unit/MASS_UNIT (→ ä 65). Endress+Hauser...
  • Page 27: Level Standard" Level Selection

    By means of the Level Mode/LEVEL_MODE parameter, select the "Pressure linearized" option. By means of the Lin. Measurand/LINEAR_ MEASURAND parameter, select the "Pressure and volume" option. Select a volume unit via the Unit Volume/ VOLUME_UNIT parameter, here m for example. Endress+Hauser...
  • Page 28 (→ ä 96), Linearization Type/L_TYPE (→ ä 97), Transducer Scale/XD_SCALE (→ ä 95) and Output Scale/OUT_SCALE (→ ä 96) parameters by means of the Analog Input Block. Result: The linearization table has been entered and the device is ready for level measurement. Endress+Hauser...
  • Page 29 3.5 m for example. See also Point 3 in the following note. P01-PMP75xxx-19-xx-xx-xx-002 By means of the Lin. Edit Mode/LINEARIZATION_ EDIT_MODE parameter, select the "Manual" option. By means of the Table Selection/LINEARIZATION_ TABLE_SELECTION parameter, select the "Edit table" option. Endress+Hauser...
  • Page 30 2nd value, and thus the highest Y-value of the linearization table, is greater than the value entered for Scale Out/SCALE_OUT, EU_100 element. The message disappears as soon as no Y_value is greater than the value for Scale Out/SCALE_OUT, EU_100 element and the table entered is active. Endress+Hauser...
  • Page 31: Height Linearized" Level Mode

    MEASURAND parameter, select the "Height + volume" option. Select the unit for the 1st measured value via the Height Unit/HEIGHT_UNIT parameter, here m for example. Select the unit for the 2nd measured variable via the Unit Volume/VOLUME_UNIT parameter, here m example. Endress+Hauser...
  • Page 32 Set the Pressure Transducer Block to the "Auto" block mode. Where necessary, configure the Channel/CHANNEL (→ ä 96), Linearization Type/L_TYPE (→ ä 97), Transducer Scale/XD_SCALE (→ ä 95) and Output Scale/OUT_SCALE (→ ä 96) parameters by means of the Analog Input Block. Endress+Hauser...
  • Page 33 The "Wet" calibration mode is not available by means of the FF operating programs. You can use the Main Line Cont./DISPLAY_MAINLINE_CONTENT parameter (→ ä 88) to specify which measured value should be displayed on the local operation. Endress+Hauser...
  • Page 34: Flow Measurement (Ff Configuration Program)

    Select a flow unit via the Unit Flow/FLOW_UNIT parameter, here m /h for example. Select the Scale Out/SCALE_OUT parameter, EU_100 element. Enter the maximum flow value of the primary device, here 6000 m /h for example. See also the layout sheet of the primary device. Endress+Hauser...
  • Page 35 (Flow-Meas. Type/FLOW_MEAS_TYPE, → ä 82). In the lower measuring range, small flow quantities (creepages) can lead to large fluctuations in the measured value. You can activate low flow cut-off via the Low Flow Cut-Off/ LOW_FLOW_CUT_OFF parameter (→ ä 84). Endress+Hauser...
  • Page 36: Totalizer

    • The parameters for the 2nd totalizer, such as Total 2. Eng. Unit/TOTALIZER_2_UNIT, Neg. Flow Tot. 2/TOTALIZER_2_MODE and Totalizer 2/TOTALIZER_2_FLOAT can be found in the Pressure Transducer Block. Endress+Hauser...
  • Page 37 Analog Input Block tranmits an alarm value to the downstream Discrete Output Block. The latter changes its output from 0 to 1 and thus resets the totalizer in the DP Flow Transducer Block to 0. The output of the Analog Input Block changes back to 0. Endress+Hauser...
  • Page 38: Parameter Description (Ff Configuration Program)

    TOTALIZER_2_VALUE. It contains all the parameters that are needed to configure this totalizer. – Diagnostic Transducer Block This Block does not return any alarm messages. It contains the simulation function for the Pressure Transducer Block, parameters to configure the alarm response and the user limits for pressure and temperature. Endress+Hauser...
  • Page 39 The guideline provides an overview of the standard function blocks that are described in FOUNDATION Fieldbus Specifications FF 890 - 894. It is designed to help operators use the blocks implemented in the Endress+Hauser field devices. Block configuration when device is delivered The block model shown below illustrates the block configuration when the device is delivered.
  • Page 40 Analog Alarm Block are not connected in the as-delivered state. (IT, DI) Note! Please note that the links between the blocks are deleted and the FF parameters are reset to the default values following a reset by means of the Restart/RESTART parameter in the Resource Block, "Default" option. Endress+Hauser...
  • Page 41: Resource Block

    Access: read only • Online: Standard block mode, the Resource Block operates in the auto mode. All the configured links between the function blocks have been established. If a link is missing, this parameter displays the "Online linking" status. Endress+Hauser...
  • Page 42 Index: 9 Data type: visible string Access: read only Manufacturer ID/ Displays the manufacturer's ID number. MANUFAC_ID Endress+Hauser: 0 x 452B48 (decimal: 4533064) Display Index: 10 Data type: unsigned16 Access: read only Device Type/DEV_TYPE Displays the device ID number.
  • Page 43 Index: 16 • Run: Standard operating mode Data type: unsigned8 • Resource: This mode is not supported by Endress+Hauser. Access: r, w • Defaults: The device data and the links of the function blocks are reset to the factory settings. The manufacturer-specific parameters of the Transducer Block are not reset to the factory settings.
  • Page 44 Enter the maximum possible number of event reports that can exist unconfirmed at the same LIM_NOTIFY time. This parameter is not evaluated by Deltabar S, Cerabar S and Deltapilot S. Entry Index: 32 Data type: unsigned8 Access: auto, OOS Endress+Hauser...
  • Page 45 – Out of service TIME_STAMP • Displays the time when the alarm occurred. SUB_CODE • Displays the reason why the alarm was reported. VALUE • Displays the value of the corresponding parameter at the time the alarm was reported. Endress+Hauser...
  • Page 46 Index: 45 The following alarms are possible: DiscAlm and BlockAlm. Data type: DS-74 UNACKNOWLEDGED Access: auto, OOS • Displays the process alarms not confirmed. UNREPORTED • Displays the process alarms not reported. DISABLED • Possibility of deactivating process alarms. Endress+Hauser...
  • Page 47 For selecting and displaying the process connection type. PROCESS_CONNEC- Options: TION_TYPE • Not used Selection • Unknown • Special Index: 47 • Oval flange Data type: unsigned16 • Thread female Access: auto, OOS • Thread male • Flange • Remote seal Endress+Hauser...
  • Page 48 • MVQ • PTFE glass • PTFE graphite • PTFE oxygen • Copper • Copper f. oxygen Factory setting: As per order specifications Up Down Feature No function. Reserved for future use. Supported/UP_DOWN_ FEATURE_SUPPORTED Index: 51 Data type:UNSIGNED8 Access: Endress+Hauser...
  • Page 49 This parameter indicates the version of the standard for electronic nameplates supported by the device. Index: 56 Data type: visible string Access: Device Tag/ The Device Tag/DEVICE_TAG currently configured via the device. DEVICE_TAG Index: 57 Data type: visible string Access: Endress+Hauser...
  • Page 50 DIRECTORY – Group ID (UINT16) – Number of the parameter in the group (UINT16) – Relative group revision index in the Resource Block of the first parameter in the group Index: 61 (UINT16) Data type: visible string Access: Endress+Hauser...
  • Page 51: Transducer Blocks

    Index: 6 display. For the Pressure, Service and Totalizer Block, see possible messages, these Operating Data type: bit string Instructions, Section 12.1 "Messages". The Display and Diagnostic Block do not display any Access: read only warnings or error messages. Endress+Hauser...
  • Page 52 A directory that specifies the number of parameter groups (data collection), and their indexes COLLECTION_ and DD item IDS, mapped in the Pressure Transducer Block. This parameter is only displayed DIRECTORY in the Pressure Transducer Block. Display Display: 0: This parameter is not used. Index: 12 Data type: Unsigned32 Access: read only Endress+Hauser...
  • Page 53 • For calibrating the lower point of the sensor characteristic curve, see the parameter description for Calibration Lowest Point/CAL_POINT_LO. Factory setting: High sensor limit (→ Sensor Range/SENSOR_RANGE, EU_100 element) Endress+Hauser...
  • Page 54 • Displays the upper measuring limit of the sensor. Index: 21 EU_0 Data type: DS-68 • Displays the lower measuring limit of the sensor. Access: read only UNITS_INDEX • Displays the unit selected. DECIMAL • Displays the number of decimal places. Endress+Hauser...
  • Page 55 Access: OOS Sensor Isolator Metal/ Displays the material of the process isolating diaphragm. SENSOR_ISOLATOR_ Display Index: 27 Data type: unsigned16 Access: read only Sensor Fill Fluid/ Displays the filling fluid. SENSOR_FILL_FLUID Display Index: 28 Data type: unsigned16 Access: read only Endress+Hauser...
  • Page 56 • Lock: Enter a number between 0 and 9999 and ≠100. • Unlock: Enter the number 100. Factory setting: Status Locking/ Displays the current locking status of the device or conditions that can lock the device STATUS_LOCKING (hardware locking, software locking). Display Index: 33 Data type:UNSIGNED 8 Endress+Hauser...
  • Page 57 • Factory setting: 0 UNITS_INDEX • Select the unit for input scaling. DECIMAL • Displays the number of decimal places. Endress+Hauser...
  • Page 58 (Primary Value) and output value of the Analog Input Block react to a change in the pressure. Index: 37 Input range: Data type: float 0.0 to 999.0 s Access: auto, OOS Factory setting: 2.0 s or as per order specification Endress+Hauser...
  • Page 59 Primary Value/PRIMARY_VALUE should be corrected. To correct the Primary Value/ PRIMARY_VALUE to 0.0 mbar, you must enter the value 2.2 here. (The following applies: PRIMARY_VALUE PRIMARY_VALUE – PRESSURE_1_INSTALL_OFFSET) – Primary Value/PRIMARY_VALUE (after entry for calib. offset) = 0.0 mbar Factory setting: Endress+Hauser...
  • Page 60 Displays the pressure that was present at the device during calibration and was used for the calibration of the upper point of the sensor characteristic curve. → See also the parameter HIGH_TRIM_MEASURE description for Calibration Highest Point/CAL_POINT_HI (→ ä 53). Display Index: 44 Data type: float Access: read only Endress+Hauser...
  • Page 61 This parameter can only be changed if the setting for the Linearization/LINEARIZATION Access: OOS parameter (→ ä 57) suits it, e.g. if the "Flow square root" option was selected for the Linearization/LINEARIZATION parameter, this parameter cannot be changed. Factory setting: Pressure and % Endress+Hauser...
  • Page 62 Data type: unsigned16 • mm Access: auto, OOS • dm • cm • m • inch • ft • User unit, → see also Customer Unit H/CUSTOMER_HEIGHT_UNIT (→ ä 63) and Cust. Unit Fact. H/CUSTOMER_UNIT_FACTOR_HEIGHT (→ ä 63) Factory setting: Endress+Hauser...
  • Page 63 – You want the measured value to be displayed in "PU" (PU: packing unit). – Primary Value/PRIMARY_VALUE = 0.5 m i 1 PU – Enter Customer Unit H/CUSTOMER_HEIGHT_UNIT: PU – Enter Cust. Unit. Fact. H/CUSTOMER_UNIT_FACTOR_HEIGHT: 2 – Result: Primary Value/PRIMARY_VALUE = 1 PU Factory setting: Endress+Hauser...
  • Page 64 – You want the measured value to be displayed in "buckets". – Primary Value/PRIMARY_VALUE = 0.01 m3 i 1 bucket – Enter Customer Unit V/CUSTOMER_UNIT_VOLUME: bucket – Enter Cust. Unit. Fact. V/CUSTOMER_UNIT_FACTOR_VOLUME : 100 – Result: Primary Value/PRIMARY_VALUE = 1 bucket Factory setting: Endress+Hauser...
  • Page 65 – You want the measured value to be displayed in "buckets". – Primary Value/PRIMARY_VALUE = 10 kg i 1 bucket – Enter Customer Unit M/CUSTOMER_UNIT_MASS: bucket – Enter Cust. Unit Fact. M/CUSTOMER_UNIT_FACTOR_MASS: 0.1 – Result: Primary Value/PRIMARY_VALUE = 1 bucket Factory setting: Endress+Hauser...
  • Page 66 • Comb. Measurand/COMBINED_MEASURAND = Height and %, Calibration Mode/CALIBRATION_MODE = Dry • Comb. Measurand/COMBINED_MEASURAND = Height and volume, Calibration Mode/CALIBRATION_MODE = Dry • Comb. Measurand/COMBINED_MEASURAND = Height and mass, Calibration Mode/CALIBRATION_MODE = Dry – Comb. Measurand = COMBINED_MEASURAND – Calibration Mode = CALIBRATION_MODE Factory setting: Endress+Hauser...
  • Page 67 "+"- or "–" key before confirming with the "E" key. This applies also if the level value is to remain unchanged. Factory setting: 100.0 Endress+Hauser...
  • Page 68 The calibration was carried out with the medium water, for example. Now the container is to Entry be used for another fluid with another density. The calibration is corrected appropriately by entering the new density value in the Process Density/PROCESS_DENSITY parameter. Index: 69 Factory setting: Data type: float Access: auto, OOS Endress+Hauser...
  • Page 69 "Lin.-Tab." window. X-Val/LINEARIZATION_ Enter the pressure value for the linearization table. → TABLE_X_VALUE See also Lin. Edit Mode/LINEARIZATION_EDIT_MODE, Line-Numb/ (manual) LINEARIZATION_TABLE_INDEX and Y-Val/LINEARIZATION_TABLE_Y_VALUE. Prerequisite: Index: 74 • Table Selection/LINEARIZATION_TABLE_SELECTION = Editor table Data type: float Access: auto, OOS Endress+Hauser...
  • Page 70 _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Entry Index: 78 Data type: visible string Access: auto, OOS Endress+Hauser...
  • Page 71 • Deltabar S = Differential • Cerabar S with gauge pressure sensors = Relative Display • Cerabar S with absolute pressure sensors = Absolute • Deltapilot S with gauge pressure sensors = Relative Index: 82 Data type: unsigned16 Access: read only Endress+Hauser...
  • Page 72 Parameter description, see → ä 74 • Level standard Once you have selected this level mode, you can use the Level Mode/LEVEL_MODE → → ä 61) to choose between "Linear", "Pressure linearized" and "Height parameter ( linearized". Factory setting: Level easy pressure Endress+Hauser...
  • Page 73 Select the density unit. The measured pressure is converted to a height using the Height Unit DENSITY_UNIT_EASY Level Easy/HEIGHT_UNIT_EASY, Density Unit Level Easy/DENSITY_UNIT_EASY and Adjust Density Level Easy/ADJUST_DENSITY_EASY parameters. Options: Index: 87 • g/cm Data type: UNSIGNED16 • kg/dm Access: OOS • kg/m • US lb/in • US lb/ft Factory setting: kg/dm³ Endress+Hauser...
  • Page 74 Full Calib. Level Easy/ Enter the value for the upper calibration point (container full). The unit is selected via the Output Unit Level Easy/OUTPUT_UNIT_EASY parameter (→ ä 73). FULL_CALIBRATION_ EASY Factory setting: 100 % Index: 93 Data type: FLOAT Access: OOS Endress+Hauser...
  • Page 75 The counter counts up to 65535 and then starts again at zero. Changes to parameters that Display configure the local operation, such as the Language/DISPLAY_LANGUAGE parameter, do not cause an increase in the counter value. Index: 14 Data type: unsigned16 Access: read only Endress+Hauser...
  • Page 76 Displays the maximum permitted absolute pressure of the sensor (resistance to overpressure). SENSOR_MAX_ ABSOLUTE_LIMIT Display Index: 21 Data type: float Access: read only Tmin Sensor/SENSOR_ Displays the lower nominal temperature limit of the sensor. TEMP_LOW_LIMIT Display Index: 22 Data type: float Access: read only Endress+Hauser...
  • Page 77 Data type: unsigned16 Access: read only Min. Meas. Press./ Displays the lowest measured pressure value (peak hold indicator). You can reset this MIN_MEASURED_ maximum indicator via the Reset Peakhold/RESET_PEAK_HOLD parameter. PRESSURE Display Index: 28 Data type: float Access: read only Endress+Hauser...
  • Page 78 Displays the number of times the specified temperature range of the electronics has been ELECTRONIC_UNDER_ undershot. TEMP_COUNTER Display Index: 35 Data type: unsigned16 Access: read only Pcb Min. Temp./ Displays the lowest electronics temperature measured. ELECTRONIC_UNDER_ TEMPERATURE Display Index: 36 Data type: float Access: read only Endress+Hauser...
  • Page 79 Displays the trend of the pressure measured value. MEASURED_VALUE_ Possibilities: increasing, decreasing, constant TREND Display Index: 40 Data type: unsigned8 Access: read only Max. Turndown/ Displays the maximum possible turndown of the transmitter. MAX_TURNDOWN Display Index: 41 Data type: float Access: read only Endress+Hauser...
  • Page 80 BA00372P (Deltapilot S), "HistoROM /M-DAT (optional)" section. Index: 48 Options: Data type: unsigned8 ® • Yes (HistoROM /M-DAT is attached to the electronic insert) Access: read only ® • No (HistoROM /M-DAT is not attached to the electronic insert) Endress+Hauser...
  • Page 81 SECONDARY_VALUE_UNIT parameter (→ ä 56) in the Pressure Transducer Block. Display Index: 52 Data type: unsigned16 Access: read only Inp. Press. Ivers/ Internal service parameter INPUT_PRESSURE_ INVERSION Index: 53 Data type: unsigned8 Access: auto, OOS (only with Service Code) Endress+Hauser...
  • Page 82 Suppressed Flow/ Displays the current flow. Depending on the flow mode selected ( Flow-Meas. Type/ SUPPRESSED_FLOW FLOW_MEAS_TYPE), a volume flow, mass flow, standard volume flow or corrected volume Display flow is displayed. Index: 15 Data type: Float Access: read only Endress+Hauser...
  • Page 83 • oz/s, oz/min • lb/s, lb/min, lb/h • ton/s, ton/min, ton/h, ton/day → • User unit, see also parameter description for Customer Unit F/CUSTOMER_UNIT_ FLOW (→ ä 84) and Cust. Unit Fact. F/CUSTOMER_UNIT_FACTOR_FLOW (→ ä 84) Factory setting: kg/s Endress+Hauser...
  • Page 84 Switching on this function stops these flow quantities from being recorded. → ä 85, Set. L. Fl. Cut-Off/SET_LOW_FLOW_CUT_OFF. Index: 19 Prerequisite: Data type: UNSIGNED8 • Deltabar S differential pressure transmitter Access: read, write, OOS Options: • Off • On Factory setting: Endress+Hauser...
  • Page 85 Access: read, write Press. Eng. Unit/ Display the pressure unit selected. PRESSURE_UNIT The pressure unit is selected by means of the Calibration Units/CAL_UNIT parameter (→ ä 54) in the Pressure Transducer Block. Display Index: 24 Data type: unsigned16 Access: read, write Endress+Hauser...
  • Page 86 Data type: unsigned8 Access: read, write Reset Totalizer 1/ You reset totalizer 1 to zero with this parameter. TOTALIZER_1_RESET Options: Selection • Abort (do not reset) • Reset Index: 29 Factory setting: Data type: unsigned8 Abort Access: read, write, OOS Endress+Hauser...
  • Page 87 CUSTOMER_UNIT_TOT_1. Selection Prerequisite: • Deltabar S differential pressure transmitter Index: 35 • Total. 2 Eng. Unit/TOTALIZER_2_UNIT = User unit Data type: visible string Factory setting: Access: read, write - - - - - - - - - - - Endress+Hauser...
  • Page 88 – Level: primary value (PV), main measured value (%), pressure, level, tank contents and Index: 13 temperature Data type: unsigned8 – Flow: primary value (PV), main measured value (%), pressure, flow, temperature, totalizer Access: auto, OOS 1 and totalizer 2 Options: • Off • On Factory setting: Endress+Hauser...
  • Page 89 • Lock: Enter a number between 0 and 9999 and 100 i. • Unlock: Enter the number 100. Factory setting: Status Locking/ Displays the current locking status of the device or conditions that can lock the device STATUS_LOCKING (hardware locking, software locking). Index: 12 Data type: unsigned8 Access: OOS Endress+Hauser...
  • Page 90 Last Diag. Code/ Displays the last error that occurred and was eliminated. LAST_DIAGNOSTIC_ Note! CODE • The message displayed here can be deleted by means of the Reset All Alarms/RESET_ Display ALL_ALARMS parameter. Index: 18 Data type: unsigned16 Access: read only Endress+Hauser...
  • Page 91 Enter the alarm response time. ALARM_DELAY Note! Entry There is no alarm if the cause of the error is eliminated within the alarm delay time. Input range: Index: 24 0 to 100 s Data type: float Access: auto, OOS Factory setting: 0.0 s Endress+Hauser...
  • Page 92 Entry Section 12.2 "Response of outputs to errors". Index: 31 Factory setting: Data type: float Upper sensor temperature operating limit +10 K (→ for the upper sensor temperature Access: auto, OOS operating limit, see Tmax Sensor/SENSOR_TEMP_HIGH_LIMIT, → ä 77) Endress+Hauser...
  • Page 93: Analog Input Block (Function Block)

    Enter the identification number for the measuring device or for each individual block. Entry The control level uses this identification number to sort alarm and event messages and initiate other processing steps. Index: 4 Input range: Data type: unsigned8 1 to 255 Access: auto, man, OOS Factory setting: Endress+Hauser...
  • Page 94 Display VALUE • Displays the process variable used for block execution Index: 7 STATUS Data type: DS-65 • Displays the status of the process variable. Access: read only Note! The unit used by the Output Scale/OUT_SCALE parameter is accepted. Endress+Hauser...
  • Page 95 • If you have selected the "Direct" option via the Linearizatin Type/L_TYPE parameter, the settings for the Transducer Scale/XD_SCALE and Output Scale/OUT_SCALE parameters must be identical. If this is not the case, the block goes to the OOS mode and the "Block config error" message is displayed in the Block Error/BLOCK_ERROR parameter. Endress+Hauser...
  • Page 96 • 6: Totalizer 1 from the DP Flow Transducer Block Factory setting: • Analog Input Block 1: Channel/CHANNEL = 1: Primary value (pressure measured value) • Analog Input Block 2: Channel/CHANNEL = 2: Secondary value (sensor temperature) • Analog Input Block 3: Channel/CHANNEL = 6: Totalizer 1 Endress+Hauser...
  • Page 97 • Displays the process variables after input scaling of the Analog Input Block. The value Index: 19 relates to a percentage of the input range Transducer Scale/XD_SCALE and is replaced by Data type: the simulation value when simulation is active. Access: read only STATUS • Displays the current status. Endress+Hauser...
  • Page 98 The following alarms are possible: HiHiAlm, HiAlm, LoLoAlm, LoAlm and BlockAlm. Data type: DS-74 UNACKNOWLEDGED Access: auto; man, OOS • Displays the process alarms not confirmed. UNREPORTED • Displays the process alarms not reported. DISABLED • Possibility of deactivating process alarms. Endress+Hauser...
  • Page 99 High High Alarm/HI_HI_ALM, High Alarm/ HI_ALM, Low Alarm/LO_ALM and Low Low Alarm/LO_LO_ALM Input range: 0.0 to 50.0 % with regard to the range of the OUT_SCALE group (→ ä 96) Factory setting: 0.5 % Endress+Hauser...
  • Page 100 • 8-15: Critical alarm with increasing priority, 8: Low priority, 15: High priority Factory setting: Low Limit/LO_LIM Enter lower limit value. Entry Input range: Range and units of Output Scale/OUT_SCALE (→ ä 96) Index: 30 Factory setting: Data type: float –INF Access: auto, man, OOS Endress+Hauser...
  • Page 101 Status display for the High Limit/HI_LIM limit value (→ ä 100). High Alarm/HI_ALM Display, selection Index: 34 Data type: DS-71 Access: auto, man, OOS Status display for the Low Limit/LO_LIM limit value (→ ä 100). Low Alarm/LO_ALM Display, selection Index: 35 Data type: DS-71 Access: auto, man, OOS Endress+Hauser...
  • Page 102 Select Alarm Mode/ Facilitates alarm mode settings for the Alarm Output Discrete/ALM_OUT_D parameter ALARM_MODE Options: • Low Cutoff/LOW_CUT Index: 43 • HiHi or LoLo Alarm activates ALARM_OUT_D/HIHI_LOLO Data type: DS66 • Hi or Lo Alarm activates ALARM_OUT_D/HI_LO Access: auto, OOS, man Endress+Hauser...
  • Page 103 Detailed description of the errors that occur within the block. BLOCK_ERR_DESC_1 Error messages: • RS_BLOCK in OOS Index: 45 • Block not scheduled Data type: UNSIGNED32 • Channel undefined Access: auto, OOS, man • L-Type undefined • AI / TRD unit inconsistent Endress+Hauser...
  • Page 104: Pressure Measurement (Via Local Operation And Fieldcare)

    – → ä 183, Table 15: OPERATING MENU → SETTINGS → EXTENDED SETUP "Pressure" – → ä 200, Table 27: OPERATING MENU → PROCESSINFO → PROCESS VALUES "Pressure". Note! • You can also specify a customer-specific unit. See parameter description for PRESS. ENG. UNIT (060) (→ ä 144). Endress+Hauser...
  • Page 105: Level Measurement (Via Local Operation And Fieldcare)

    % must not be in direct proportion to the measured pressure. A linearization table must be entered for the 2nd measured variable. The 2nd measured variable is assigned to the 1st measured variable by means of this table. Endress+Hauser...
  • Page 106: Level Easy Pressure" Level Selection

    LEVEL SELECTION parameter. See Table, Step 9. See Table, Step 10. Local operation: Menu path: GROUP SELECTION → MEASURING MODE "Level" → LEVEL SELECTION FieldCare: Menu path: OPERATING MENU → SETTINGS → BASIC SETUP → MEASURING MODE "Level" → LEVEL SELECTION Endress+Hauser...
  • Page 107 If operating via the local operation, the EMPTY CALIB. (010) (→ ä 149) and FULL CALIB. (004) (→ ä 149) parameters also show the pressure present at the device. For operation using FieldCare, the pressure present at the device is displayed in the PROCESS VALUES group. Endress+Hauser...
  • Page 108 LEVEL SELECTION See Table, Step 10. See Table, Step 11. Local operation: See Table, Step 12. Select the BASIC SETUP function group. Menu path: See Table, Step 13. GROUP SELECTION → OPERATING MENU → SETTINGS → BASIC SETUP Endress+Hauser...
  • Page 109 The measuring range is set for 0 to 1000 l (264 US gal). Note! For this level mode, the measured variables %, level, volume and mass are available. → See also parameter description for OUTPUT UNIT (023), → ä 148. Endress+Hauser...
  • Page 110: Level Easy Height" Level Selection

    Menu path: GROUP SELECTION → MEASURING MODE FieldCare: Menu path: OPERATING MENU → SETTINGS → P01-PMC71xxx-19-xx-xx-xx-009 BASIC SETUP → MEASURING MODE Fig. 23: Calibration with reference pressure – wet calibration See Table, Steps 10 and 11. See Table, Step 12. See Table, Step 13. Endress+Hauser...
  • Page 111 The measuring range is set for 0 to 1000 l (264 US gal). Note! For this level mode, the measured variables %, level, volume and mass are available. → See also parameter description for OUTPUT UNIT (023), → ä 152. Endress+Hauser...
  • Page 112 LEVEL SELECTION See Table, Step 11. See Table, Steps 13 and 14. Local operation: See Table, Steps 15 and 16. Select the BASIC SETUP function group. Menu path: GROUP SELECTION → OPERATING MENU → SETTINGS → BASIC SETUP Endress+Hauser...
  • Page 113 See Table, Step 11. See Table, Step 12. See Table, Step 13. Note! For this level mode, the measured variables %, level, volume and mass are available. → See also parameter description for OUTPUT UNIT (023), → ä 152. Endress+Hauser...
  • Page 114: Level Standard" Level Selection, "Linear" Level Mode

    If necessary, select the "Level standard" level mode using the LEVEL SELECTION parameter. See Table, Step 11. See Table, Step 12. Local operation: Menu path: GROUP SELECTION → MEASURING MODE FieldCare: Menu path: OPERATING MENU → SETTINGS → BASIC SETUP → MEASURING MODE Endress+Hauser...
  • Page 115 For this level mode, the measured variables %, level, volume and mass are available. → ä 158 The EMPTY PRESSURE (710) (→ ä 161) and FULL PRESSURE (711) (→ ä 161) parameters display the pressure values belonging to the EMPTY CALIB. and FULL CALIB. parameters. Endress+Hauser...
  • Page 116 BASIC SETUP → MEASURING MODE ➃ ➀ –0.5 m r = 1 kg P01-PMP75xxx-19-xx-xx-xx-003 Fig. 28: Calibration without reference pressure – dry calibration See Table, Step 9. See Table, Step 10. See Table, Step 11. See Table, Step 12. Endress+Hauser...
  • Page 117 For this level mode, the measured variables %, level, volume and mass are available. → ä 158 You can also specify customer-specific units. See parameter descriptions for PRESS. ENG. UNIT (060) (→ ä 155), HEIGHT UNIT (708) (→ ä 158), UNIT VOLUME (313) (→ ä 159) and MASS UNIT (709) (→ ä 160). Endress+Hauser...
  • Page 118: Level Standard" Level Selection

    If necessary, select the "Level standard" level mode using the LEVEL SELECTION parameter. Local operation: Menu path: GROUP SELECTION → MEASURING MODE "Level" → LEVEL SELECTION FieldCare: Menu path: OPERATING MENU → SETTINGS → BASIC SETUP → MEASURING MODE "Level" → LEVEL SELECTION Endress+Hauser...
  • Page 119 LINE-NUMB: enter the value of the corresponding point. X-VAL.: The hydrostatic pressure present is displayed. The X-VAL. displayed is saved by confirming the Y- value. See following line, Y-VAL. Y-VAL.: Enter the volume value, here 0 m for example, and confirm the value. Endress+Hauser...
  • Page 120 Y-VALUE of the linearization table, is greater than the value entered for TANK CONTENT MAX. The message disappears as soon as no Y-VALUE is greater than the value for TANK CONTENT MAX. and the table entered is active. Endress+Hauser...
  • Page 121 LINEARIZATION – FieldCare • For a description of further relevant parameters, see – → ä 183, Table 16: OPERATING MENU → SETTINGS → EXTENDED SETUP "Level" – → ä 201, Table 28: OPERATING MENU → PROCESSINFO → PROCESS VALUES "Level". Endress+Hauser...
  • Page 122 Input Block, see Page 93 parameter descriptions. For this purpose, equate XD SCALE and OUT SCALE or the OUT value with the measured value. Result: The linearization table has been entered and the device is ready for level measurement. Endress+Hauser...
  • Page 123 (→ MINIMUM SPAN (591), → ä 198). The message disappears as soon as the highest X-VALUE is greater than the minimum span and the table entered is active. Endress+Hauser...
  • Page 124: Height Linearized" Level Mode

    If necessary, select the "Level standard" level mode using the LEVEL SELECTION parameter. Local operation: Menu path: GROUP SELECTION → MEASURING MODE "Level" → LEVEL SELECTION FieldCare: Menu path: OPERATING MENU → SETTINGS → BASIC SETUP → MEASURING MODE "Level" → LEVEL SELECTION Endress+Hauser...
  • Page 125 Menu path: OPERATING MENU → SETTINGS → LINEARIZATION Select the TANK CONTENT MIN parameter. Specify the minimum tank contents to be expected, here for example. Select the TANK CONTENT MAX parameter . Specify the maximum tank contents to be expected, here for example. Endress+Hauser...
  • Page 126 X-VALUE is greater than the maximum level and the table entered is active. You can use the MENU DESCRIPTOR (416) parameter (→ ä 194) to specify which measured value should be displayed on the local operation. Endress+Hauser...
  • Page 127 The calibration for the 1st measured variable is carried out. Perform linearization (calibration for the 2nd measured variable) Change the function group. Local operation: Menu path: GROUP SELECTION →OPERATING MENU → SETTINGS → LINEARIZATION FieldCare: Menu path: OPERATING MENU → SETTINGS → LINEARIZATION Endress+Hauser...
  • Page 128 UNIT (060) (→ ä 155), HEIGHT UNIT (708) (→ ä 170), UNIT VOLUME (313) (→ ä 171) and MASS UNIT (709) (→ ä 172). You can use the MENU DESCRIPTOR (416) parameter (→ ä 194) to specify which measured value should be displayed on the local operation. Endress+Hauser...
  • Page 129: Flow Measurement (Via Local Operation And Fieldcare)

    /h for example. See also the layout sheet of the primary device. Select the MAX PRESS. FLOW parameter. Enter the maximum pressure, here 400 mbar (6 psi) for example. See also the layout sheet of the primary device. Result: The device is configured for flow measurement. Endress+Hauser...
  • Page 130 MASS FLOW UNIT (571) (→ ä 181). In the lower measuring range, small flow quantities (creepages) can lead to large fluctuations in the measured value. You can activate low flow cut-off via the LOW FLOW CUT-OFF (442) parameter (→ ä 185). Endress+Hauser...
  • Page 131: Totalizer

    The TOTALIZER 2 and TOTAL. 2 OVERFLOW parameters display the totalized flow value of the second totalizer. → ä 202 ff, PROCESS VALUES function group. • You can use the MENU DESCRIPTOR (416) parameter (→ ä 194) to specify which measured value should be displayed on the local operation. Endress+Hauser...
  • Page 132 Analog Input Block tranmits an alarm value to the downstream Discrete Output Block. The latter changes its output from 0 to 1 and thus resets the totalizer in the DP Flow Transducer Block to 0. The output of the Analog Input Block changes back to 0. Endress+Hauser...
  • Page 133: Parameter Description (Local Operation And Fieldcare)

    • Select the menu language for FieldCare using the "Language button" in the configuration window. Select the menu language for the FieldCare frame via the "Extra" menu → "Options" → "Display" → "Language". Options: • Deutsch • English • Français • Español • Chinese (CHS) • Japanese (JPN) Factory setting: English Endress+Hauser...
  • Page 134 • The MEASURING MODE parameter is displayed in FieldCare in the QUICK SETUP menus and in the BASIC SETUP function group (OPERATING MENU → SETTINGS → BASIC SETUP). Options: • Pressure • Level • Deltabar S: Flow Factory setting: • Cerabar S and Deltabar S: Pressure • Deltapilot S: Level Endress+Hauser...
  • Page 135 → For LEVEL SELECTION = "Level easy height" → ä 150, Table 9: OPERATING MENU → SETTINGS → BASIC SETUP "Level", LEVEL SELECTION "Level easy height". → For LEVEL SELECTION = "Level standard" → ä 155, Table 10: OPERATING MENU → SETTINGS → BASIC SETUP "Level", LEVEL SELECTION "Level standard". Endress+Hauser...
  • Page 136 MEASURED VALUE. The device has to be recalibrated if the measuring mode is changed. → ä 104, Section 8 "Pressure measurement (via local operation and FieldCare)". Options: • Pressure • Level • Deltabar S: Flow Factory setting: • Cerabar S and Deltabar S: Pressure • Deltapilot S: Level Endress+Hauser...
  • Page 137 The damping affects the speed at which all subsequent elements, such as the local operation, measured value and OUT value of the Analog Input Block react to a change in the pressure. Input range: 0.0 to 999.0 s Factory setting: 2.0 s or as per order specifications Endress+Hauser...
  • Page 138 • Pressure • Level • Deltabar S: Flow Factory setting: • Cerabar S and Deltabar S: Pressure • Deltapilot S: Level LEVEL SELECTION (020) Select the level mode. →Parameter description, → ä 135. Selection Factory setting: Level easy pressure Endress+Hauser...
  • Page 139 "+"- or "–" key before confirming with the "E" key. This applies also if the level value is to remain unchanged. Factory setting: Endress+Hauser...
  • Page 140 – → ä 185 ff, Table 17: OPERATING MENU → SETTINGS → EXTENDED SETUP "Flow" – → ä 192 ff, Table 20: OPERATING MENU → SETTINGS → TOTALIZER SETUP – → ä 129 ff, Section 10 "Flow measurement (via local operation and FieldCare)". Endress+Hauser...
  • Page 141 The damping affects the speed at which all subsequent elements, such as the local operation, measured value and OUT value of the Analog Input Block react to a change in the pressure. Input range: 0.0 to 999.0 s Factory setting: 2.0 s or as per order specifications Endress+Hauser...
  • Page 142 "Confirm" option. This means that you are assigning the value 0.0 to the pressure present. – MEASURED VALUE (after pos. zero adjust) = 0.0 mbar The CALIB. OFFSET parameter displays the resulting pressure difference (offset) by which the MEASURED VALUE was corrected. Options: • Abort • Confirm Factory setting: Endress+Hauser...
  • Page 143 Parameters marked with an asterisk "*" CUSTOMER UNIT P are not always displayed. The display depends on the settings of other parameters, see the following Table. CUST. UNIT FACT. P DAMPING VALUE P01-xMx7xxxx-19-xx-xx-en-004 Fig. 41: BASIC SETUP function group for the "Pressure" measuring mode Endress+Hauser...
  • Page 144 The maximum number of characters in the counter is again limited to five. For example if the customer-specific unit "crates/m2" was selected, "crate/m2" would be displayed. FieldCare accepts units with eight characters at most. The slash counts as one character. Factory setting: _ _ _ _ _ _ _ _ Endress+Hauser...
  • Page 145 The damping affects the speed at which all subsequent elements, such as the local operation, measured value and OUT value of the Analog Input Block react to a change in the pressure. Input range: 0.0 to 999.0 s Factory setting: 2.0 s or as per order specifications Endress+Hauser...
  • Page 146 The following parameters are displayed if you have selected the "Level easy pressure" option for the LEVEL SELECTION parameter. Specify two pressure-level value pairs for this level mode. The two calibration modes, "Wet" and "Dry", are available. Prerequisite: • MEASURING MODE = Level (→ ä 134). • LEVEL SELECTION = Level easy pressure (→ ä 135.) Endress+Hauser...
  • Page 147 – You want the measured value to be displayed in "PU" (PU: packing unit). – MEASURED VALUE =10000 Pa i 1 PU – Entry CUSTOMER UNIT P: PU – Entry CUST. UNIT FACT. P: 0.0001 – Result: MEASURED VALUE = 1 PU Factory setting: Endress+Hauser...
  • Page 148 "+"- or "–" key before confirming with the "E" key. This applies also if the level value is to remain unchanged. Factory setting: Endress+Hauser...
  • Page 149 The damping affects the speed at which all subsequent elements, such as the local operation, measured value and OUT value of the Analog Input Block react to a change in the pressure. Input range: 0.0 to 999.0 s Factory setting: 2.0 s or as per order specifications Endress+Hauser...
  • Page 150 The two calibration modes, "Wet" and "Dry", are available. Prerequisite: • MEASURING MODE = Level (→ ä 134). • LEVEL SELECTION = Level easy height (→ ä 135.) Endress+Hauser...
  • Page 151 – You want the measured value to be displayed in "PU" (PU: packing unit). – MEASURED VALUE =10000 Pa i 1 PU – Entry CUSTOMER UNIT P: PU – Entry CUST. UNIT FACT. P: 0.0001 – Result: MEASURED VALUE = 1 PU Factory setting: Endress+Hauser...
  • Page 152 • US lb/in • US lb/ft Factory setting: kg/dm ADJUST DENSITY (007) Enter the density of the fluid. The measured pressure is converted to a height using the Entry HEIGHT UNIT, DENSITY UNIT and ADJUST DENSITY parameters. Factory setting: Endress+Hauser...
  • Page 153 The values entered for the FULL CALIB. and FULL HEIGHT parameters form the height- level value pair for the upper calibration point. The unit is selected via the OUTPUT UNIT parameter (→ ä 152). Prerequisite: • CALIBRATION MODE = Dry Factory setting: 100.0 Endress+Hauser...
  • Page 154 → ä 165, → Fig. 47 BASIC SETUP function group for the "Level" measuring mode and "Pressure linearized" level mode → ä 169, → Fig. 48 BASIC SETUP function group for the "Level" measuring mode and "Height linearized" level mode. Endress+Hauser...
  • Page 155 MEASURED VALUE no longer display the same value. → See also parameter descriptions for XD SCALE, OUT SCALE and OUT value. Factory setting: Depends on the sensor nominal measuring range mbar or bar or as per order specifications Endress+Hauser...
  • Page 156 → For LEVEL MODE = Pressure linearized, → ä 166 ff, Table 12: OPERATING MENU → SETTINGS → BASIC SETUP "Level", LEVEL MODE "Pressure linearized". → For LEVEL MODE = Height linearized, → ä 170 ff, Table 13: OPERATING MENU → SETTINGS → BASIC SETUP "Level", LEVEL MODE "Height linearized". Endress+Hauser...
  • Page 157 CUSTOMER UNIT H depends on the settings of other parameters, see the following Table. CUST. UNIT FACT. H 100% POINT ZERO OFFSET DAMPING VALUE P01-xMx7xxxx-19-xx-xx-en-006 Fig. 45: BASIC SETUP function group for the "Level" measuring mode and "Linear" level mode Endress+Hauser...
  • Page 158 The maximum number of characters in the counter is again limited to five. For example if the customer-specific unit "crates/m2" was selected, "crate/m2" would be displayed. FieldCare accepts units with eight characters at most. The slash counts as one character. Factory setting: _ _ _ _ _ _ _ _ Endress+Hauser...
  • Page 159 The maximum number of characters in the counter is again limited to five. For example if the customer-specific unit "crates/m2" was selected, "crate/m2" would be displayed. FieldCare accepts units with eight characters at most. The slash counts as one character. Factory setting: _ _ _ _ _ _ _ _ Endress+Hauser...
  • Page 160 – You want the measured value to be displayed in "buckets". – MEASURED VALUE = 10 kg i 1 bucket – Entry CUSTOMER UNIT M: bucket – Entry CUST. UNIT FACT. M: 0.1 – Result: MEASURED VALUE = 1 bucket Factory setting: Endress+Hauser...
  • Page 161 • CALIBRATION MODE = Wet Factory setting: High sensor limit (→ see PRESS. SENS HILIM, → ä 198) ADJUSTED DENSITY (810) Displays the density calculated from the upper and lower level point. Display Prerequisite: • CALIBRATION MODE = Wet, LIN. MEASURAND = Level Endress+Hauser...
  • Page 162 The maximum number of characters in the counter is again limited to five. For example if the customer-specific unit "crates/m2" was selected, "crate/m2" would be displayed. FieldCare accepts units with eight characters at most. The slash counts as one character. Factory setting: _ _ _ _ _ _ _ _ Endress+Hauser...
  • Page 163 The maximum number of characters in the counter is again limited to five. For example if the customer-specific unit "crates/m2" was selected, "crate/m2" would be displayed. FieldCare accepts units with eight characters at most. The slash counts as one character. Factory setting: _ _ _ _ _ _ _ _ Endress+Hauser...
  • Page 164 Zero point shift Device is mounted above the level lower-range value: a positive value has to be entered for ZERO POSITION. Device is mounted below the level lower-range value: a negative value has to be entered for ZERO POSITION. Endress+Hauser...
  • Page 165 CUST. UNIT FACT. M HYDR. PRESS MIN. HYDR. PRESS MAX. DAMPING VALUE P01-xMx7xxxx-19-xx-xx-en-007 Fig. 47: BASIC SETUP function group for the "Level" measuring mode and "Pressure linearized" level mode, continue calibration with LINEARIZATION function group → ä 186 ff. Endress+Hauser...
  • Page 166 Following a change in the unit, the digital output value of the Analog Input value OUT and the local operation or the MEASURED VALUE no longer display the same value. → See also parameter descriptions for XD SCALE, OUT SCALE and OUT value. Factory setting: Endress+Hauser...
  • Page 167 Following a change in the unit, the digital output value of the Analog Input value OUT and the local operation or the MEASURED VALUE no longer display the same value. → See also parameter descriptions for XD SCALE, OUT SCALE and OUT value. Factory setting: Endress+Hauser...
  • Page 168 The damping affects the speed at which all subsequent elements, such as the local operation, measured value and OUT value of the Analog Input Block react to a change in the pressure. Input range: 0.0 to 999.0 s Factory setting: 2.0 s or as per order specifications Endress+Hauser...
  • Page 169 CUST. UNIT FACT. H DENSITY UNIT 100% POINT ZERO POSITION DAMPING VALUE P01-xMx7xxxx-19-xx-xx-en-009 Fig. 48: BASIC SETUP function group for the "Level" measuring mode and "Height linearized" level mode, continue calibration with LINEARIZATION function group → ä 186 ff. Endress+Hauser...
  • Page 170 • COMB. MEASURAND = Height and volume, height and mass or height and % Options: • mm • dm • cm • m • inch • ft • User unit, → see also the following parameter descriptions for CUSTOMER UNIT H and CUST. UNIT. FACT. H. Factory setting: Endress+Hauser...
  • Page 171 Following a change in the unit, the digital output value of the Analog Input value OUT and the local operation or the MEASURED VALUE no longer display the same value. → See also parameter descriptions for XD SCALE, OUT SCALE and OUT value. Factory setting: Endress+Hauser...
  • Page 172 Following a change in the unit, the digital output value of the Analog Input value OUT and the local operation or the MEASURED VALUE no longer display the same value. → See also parameter descriptions for XD SCALE, OUT SCALE and OUT value. Factory setting: Endress+Hauser...
  • Page 173 Enter the maximum level to be expected. Entry The input limits for the calibration (editing limits) are derived from the value entered. The closer the value entered corresponds to the maximum level to be expected, the more accurate the measurement result. Factory setting: 100.0 Endress+Hauser...
  • Page 174 Displays the density calculated from the upper and lower level point. Display Prerequisite: • COMB. MEASURAND = Height and volume, CALIBRATION MODE = Wet • COMB. MEASURAND = Height and mass, CALIBRATION MODE = Wet • COMB. MEASURAND = Height and %, CALIBRATION MODE = Wet Endress+Hauser...
  • Page 175 • COMB. MEASURAND = %-height and %, CALIBRATION MODE = Dry Options: • mm • dm • cm • m • inch • ft • User unit, → see also the following parameter descriptions for CUSTOMER UNIT H and CUST. UNIT. FACT. H. Factory setting: Endress+Hauser...
  • Page 176 • COMB. MEASURAND = %-height and %, CALIBRATION MODE = Dry Example: – The 100 %-point should correspond to 4 m (13 ft). – Select the "m" unit via the HEIGHT UNIT parameter. – Enter the value "4" for this parameter (100% POINT). Factory setting: Endress+Hauser...
  • Page 177 The damping affects the speed at which all subsequent elements, such as the local operation, measured value and OUT value of the Analog Input Block react to a change in the pressure. Input range: 0.0 to 999.0 s Factory setting: 2.0 s or as per order specifications Endress+Hauser...
  • Page 178 – → ä 192, Table 20: OPERATING MENU → SETTINGS → TOTALIZER SETUP – → ä 202, Table 29: OPERATING MENU → PROCESSINFO → PROCESS VALUES "Flow" – → ä 129 ff, Section 10 "Flow measurement (via local operation and FieldCare)". Endress+Hauser...
  • Page 179 The maximum number of characters in the counter is again limited to five. For example if the customer-specific unit "crates/m2" was selected, "crate/m2" would be displayed. FieldCare accepts units with eight characters at most. The slash counts as one character. Factory setting: _ _ _ _ _ _ _ _ Endress+Hauser...
  • Page 180 Following a change in the unit, the digital output value of the Analog Input value OUT and the local operation or the MEASURED VALUE no longer display the same value. → See also parameter descriptions for XD SCALE, OUT SCALE and OUT value. Factory setting: Endress+Hauser...
  • Page 181 Following a change in the unit, the digital output value of the Analog Input value OUT and the local operation or the MEASURED VALUE no longer display the same value. → See also parameter descriptions for XD SCALE, OUT SCALE and OUT value. Factory setting: kg/s Endress+Hauser...
  • Page 182 The damping affects the speed at which all subsequent elements, such as the local operation, measured value and OUT value of the Analog Input Block react to a change in the pressure. Input range: 0.0 to 999.0 s Factory setting: 2.0 s or as per order specifications Endress+Hauser...
  • Page 183 Factory setting: °C Table 16: OPERATING MENU → SETTINGS → EXTENDED SETUP "Level" Parameter name Description Prerequisite: • MEASURING MODE = Level Note: • See also → ä 105 ff, Section 9 "Level measurement (via local operation and FieldCare)". Endress+Hauser...
  • Page 184 DENSITY parameters must be entered correctly before changing the calibration mode. If the pressure falls with increasing levels (LIN. MEASURED: volume), such as in the case of a residual volume measurement, a negative value must be entered for this parameter. Factory setting: Endress+Hauser...
  • Page 185 → See also LOW FLOW CUT-OFF. Prerequisite: • LOW FLOW CUT-OFF = On Input range: Switch-off point: 0 to 50 % of end flow value (→ MAX. FLOW). ➀ ➁ Δp Δp P01-PMD7xxxx-05-xx-xx-xx-000 Factory setting: 5 % (of end flow value) Endress+Hauser...
  • Page 186 To make sure measuring also runs properly when entering a new table, there is another table, the editor table, for entering new values. Options: • View meas. table • Editor table Factory setting: View meas. table Endress+Hauser...
  • Page 187 – X-VAL.: The hydrostatic pressure present is measured. Depending on the setting in the COMB. MEASURAND parameter, the measured pressure is converted to a level unit or a % and displayed. The value is saved by confirming the Y-value. Factory setting: LINE-NUMB = 1, X-VAL. = 0.0, Y-VAL. = 0.0 Endress+Hauser...
  • Page 188 Enter tank description. (max. 32 alphanumeric characters) Entry Factory setting: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Endress+Hauser...
  • Page 189 To make sure measuring also runs properly when entering a new table, there is another table, the editor table, for entering new values. Options: • View meas. table • Editor table Factory setting: View meas. table Endress+Hauser...
  • Page 190 In the "Semiautomatic" entry mode, the container is filled or emptied in stages. The entry mode) X-VAL. displays the measured hydrostatic pressure. Display Prerequisite: • TABLE SELECTION = Editor table FieldCare The X-VAL. is saved by confirming the Y-value. → See also LIN. EDIT MODE, LINE-NUMB and Y-VAL. Endress+Hauser...
  • Page 191 Enter tank description. (max. 32 alphanumeric characters) Entry Factory setting: _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Endress+Hauser...
  • Page 192 – Index 102: TOTALIZER 1 UNIT (general) – Index 156 (398): FLOW-MEAS. TYPE "Volume operat. cond." – Index 168 (662): FLOW-MEAS. TYPE "Mass" – Index 170 (664): FLOW-MEAS. TYPE "Gas. std. cond." – Index 172 (666): FLOW-MEAS. TYPE "Gas. norm cond." Factory setting: Endress+Hauser...
  • Page 193 – Index 103: TOTALIZER 2 UNIT (general) – Index 157 (399): FLOW-MEAS. TYPE "Volume operat. cond." – Index 169 (663): FLOW-MEAS. TYPE "Mass" – Index 171 (665): FLOW-MEAS. TYPE "Gas. std. cond." – Index 173 (667): FLOW-MEAS. TYPE "Gas. norm cond." Factory setting: Endress+Hauser...
  • Page 194 • Primary value (PV) • Main measured value (%) • Pressure • Flow • Level • Tank content • Temperature • Error number • Totalizer 1 • Totalizer 2 The selection depends on the measuring mode chosen. Factory setting: Primary value (PV) Endress+Hauser...
  • Page 195 → See also Operating Instructions BA00301P (Deltabar S) and BA00302P (Cerabar S) or BA00372P (Deltapilot S), "Function of the operating keys" section. Input range: 4 to 13, 4: contrast weaker (brighter), 13: contrast stronger (darker). Factory setting: Endress+Hauser...
  • Page 196 Set tag number for the device. DEVICE DESIGN. (350) Displays the device designation and the order code. Display HARDWARE REV. (266) Displays the revision number of the main electronics Display e.g. V02.00.00 SOFTWARE VERSION (264) Displays the software version Display e.g.: V 03.00.00 Endress+Hauser...
  • Page 197 BA00301P), Cerabar S (BA00302P) or Deltapilot S (BA00372P), "Nameplates" section) PROC. CONN. TYPE (482) For selecting and displaying the process connection type. Selection Options: • Not used • Unknown • Special • Oval flange • Thread female • Thread male • Flange • Remote seal Endress+Hauser...
  • Page 198 Displays the serial number of the sensor (11 alphanumeric characters). Display PRESS. SENS LOLIM (484) Displays the lower measuring limit of the sensor. Display PRESS. SENS HILIM (485) Displays the upper measuring limit of the sensor. Display MINIMUM SPAN (591) Displays the smallest possible span. Display Endress+Hauser...
  • Page 199 Displays the revision or version of a complete device (HW+SW) within a device type. Display DD REVISION (985) Displays the DD version initially certified. Display DEVICE ADDRESS (984) Displays the device address currently configured and valid. Display The factory setting is 247. Endress+Hauser...
  • Page 200 Table 27: OPERATING MENU → PROCESSINFO → PROCESS VALUES "Pressure" Parameter name Description Prerequisite: • MEASURING MODE = Pressure MEASURED VALUE (679) Displays the measured value Display In the "Pressure" measuring mode, this value corresponds to the PRESSURE parameter. Endress+Hauser...
  • Page 201 SENSOR TEMP. (367) Displays the temperature currently measured in the sensor. This can deviate from the Display process temperature. MEAS. VAL. TREND (378) Displays the trend of the primary value of the Transducer Block. Display Possibilities: increasing, decreasing, constant Endress+Hauser...
  • Page 202 Displays the overflow value of totalizer 1. → See also TOTALIZER 1. Display TOTALIZER 2 (657) Displays the total flow value of totalizer 2. You cannot reset totalizer 2. The TOTAL. 2 Display OVERFLOW parameter displays the overflow. → See also example for TOTALIZER 1. Endress+Hauser...
  • Page 203 Displays the number of times the specified temperature range of the electronics has been (488) overshot. Display PCB MAX. TEMP. (490) Displays the highest electronics temperature measured. Display PCB COUNT:T < Tmin (492) Displays the number of times the specified temperature range of the electronics has been Display undershot. Endress+Hauser...
  • Page 204 SIMULATION function group, → ä 206, Table 33: OPERATING MENU → DIAGNOSTICS → SIMULATION for the MESSAGES function group, → ä 207, Table 34: OPERATING MENU → DIAGNOSTICS → MESSAGES for the USER LIMITS function group, → ä 208, Table 35: OPERATING MENU → DIAGNOSTICS → USER LIMITS Endress+Hauser...
  • Page 205 ® • A HistoROM /M-DAT is attached to the electronic insert (HistoROM AVAIL. = Yes) Options: • Abort • HistoROM → Device • Device → HistoROM Factory setting: ® Abort (if HistoROM /M-DAT is attached to the electronic insert) Endress+Hauser...
  • Page 206 MODE = Tank content SIM. ERROR NO. (476) Enter the message number. → See also SIMULATION MODE. Entry → See also these Operating Instructions, Section 12.1 "Messages", "Code" table column. Prerequisite: • SIMULATION MODE = Alarm/warning Factory setting: 613 (simulation active) Endress+Hauser...
  • Page 207 (A) or as in the event of a warning (W). Enter the corresponding message number for this parameter. → See also SELECT ALARMTYPE. → See also these Operating Instructions, Section 12.1 "Messages" and Section 12.2 "Response of outputs to errors". Endress+Hauser...
  • Page 208 → See also these Operating Instructions, Section 12.1 "Messages" Table, code E731 and Section 12.2 "Response of outputs to errors". Factory setting: High sensor limit • 1.1 (→ For the high sensor limit, see PRESS. SENS HILIM.) Endress+Hauser...
  • Page 209: Troubleshooting

    – The Alarm Status/ALARM_STATUS parameter shows all the messages present in descending order of priority. You can scroll through all the messages pending using the S or O key. • FieldCare: The Alarm Status/ALARM_STATUS parameter shows the message with the highest priority. → See the "Priority" column. Endress+Hauser...
  • Page 210 • If the device detects a defect in the local operation during initialization, special error messages are generated. → For the error messages, → ä 218, Section 12.1.1 "Local operation error messages". • For support and further information, please contact Endress+Hauser Service. Code Error...
  • Page 211 – Add to linearization table. invalid. Min. 2 points of less than 2 points. Accept linearization table again if necessary. 613 (W613) Warning I>Simulation is active – Simulation is switched on, i.e. – Switch off simulation. the device is not measuring at present. Endress+Hauser...
  • Page 212 – Device reset codes such as 7864 (onsite) or 40864 (bus) do not have any effect on the HistoROM. That means that if you perform a reset, the configurations in the HistoROM and in the device may not be the same. Endress+Hauser...
  • Page 213 – The temperature measured in – Increase ambient temperature. temperature the electronics is smaller than Insulate device if necessary. the lower nominal temperature of the electronics (–43 °C (-45 °F)). – Unsuitable download carried – Check configuration and out. perform download again. Endress+Hauser...
  • Page 214 TI00382P (Deltabar S), TI00383P (Cerabar S) or TI00416P (Deltapilot S). – Process temperature is outside – Check temperature present, permitted range. reduce or increase if necessary. – Sensor defect. – If the process temperature is within the permitted range, replace sensor. Endress+Hauser...
  • Page 215 – Briefly disconnect device from the power supply. – Main electronics defect. – Replace main electronics. 739 (A739) Alarm B>Measurement error – Fault in the main electronics. – Briefly disconnect device from the power supply. – Main electronics defect. – Replace main electronics. Endress+Hauser...
  • Page 216 TI00383P (Cerabar S) or TI00416P (Deltapilot S). – Main electronics defect. – Replace main electronics. 745 (W745) Warning C>Sensor data unknown 17 – Sensor does not suit the device – Replace sensor with a suitable (electronic sensor nameplate). sensor. Device continues measuring. Endress+Hauser...
  • Page 217 – Perform reset (code 7864) and configuration of the device that recalibrate the device. do not suit one another. e.g. if the "1 (linearization table)" option was selected for Linearization/LINEARIZATION and the "1347 (m /s)" unit for Primary Value Type/PRIMARY_ VALUE_UNIT. Endress+Hauser...
  • Page 218 If the device detects a defect in the local operation during initialization, the following error messages can be displayed: Message Measure Initialization, VU Electr. Defect A110 Exchange local operation. Initialization, VU Electr. Defect A114 Initialization, VU Electr. Defect A281 Initialization, VU Checksum Err. A110 Initialization, VU Checksum Err. A112 Initialization, VU Checksum Err. A171 Endress+Hauser...
  • Page 219: Response Of Outputs To Errors

    122 for "Sensor displays a 3-digit number such as 613 for digit number such as 731 for "Pmax connection error, incorrect data." "Simulation is active". ALARM WINDOW undershot". FF configuration program: Diagnostic Transducer Block. Menu path FieldCare: OPERATING MENU → MESSAGES Endress+Hauser...
  • Page 220: Confirming Messages

    FF configuration program: The parameters are in the Diagnostic Transducer Blocks. FieldCare: Menu path for Alarm Displ. Time/ALARM_DISPLAY_TIME and Ack. Alarm Mode/ACKNOW- LEDGE_ALARM_MODE: OPERATING MENU → DIAGNOSTICS → MESSAGES Endress+Hauser...
  • Page 221: Appendix

    "User unit" option was selected for the Press. Eng. Unit/ parameter. These parameters are indicated with a "*". • For a description of the parameters, see Section 11 "Parameter description (Local operation and FieldCare)". The exact dependency of individual parameters on one another is explained here. Endress+Hauser...
  • Page 222 Appendix Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus Endress+Hauser...
  • Page 223 There are parameters that are only displayed if other parameters are appropriately configured. For example the Customer Unit P/CUSTOMER_UNIT_PRESSURE parameter is only displayed if the "User unit" option was selected for the Press. Eng. Unit/PRESSURE UNIT parameter. These parameters are indicated with a "*". P01-xxxxxxxx-19-xx-xx-en-167 Endress+Hauser...
  • Page 224 There are parameters that are only displayed if other parameters are appropriately configured. For example the Customer Unit P/CUSTOMER_UNIT_PRESSURE parameter is only displayed if the "User unit" option was selected for the Press. Eng. Unit/PRESSURE_UNIT parameter. These parameters are indicated with a "*". P01-xxxxxxxx-19-xx-xx-en-149 Endress+Hauser...
  • Page 225 There are parameters that are only displayed if other parameters are appropriately configured. For example the Customer Unit H/CUSTOMER_HEIGHT_UNIT parameter is only displayed if the "User unit" option was selected for the Height Unit/HEIGHT_UNIT parameter. These parameters are indicated with a "*". P01-xxxxxxxx-19-xx-xx-en-150 Endress+Hauser...
  • Page 226 There are parameters that are only displayed if other parameters are appropriately configured. For example the Tot. 1 User Unit/CUSTOMER_UNIT_TOT_1 parameter is only displayed if the "User unit" option was selected for the Totalizer 1 Unit/TOTALIZER_1_UNIT parameter. These parameters are indicated with a "*". P01-xxxxxxxx-19-xx-xx-en-151 Endress+Hauser...
  • Page 227 DEVICE DESIGN. HARDWARE REV. MAT. MEMBRANE SOFTWARE VERSION FILLING FLUID CONFIG RECORDER Tmin SENSOR PCB TEMPERATURE Tmax SENSOR ALLOWED MIN. TEMP SENSOR H/WARE REV. ALLOWED MAX. TEMP DIP STATUS 1) Display via on-site display only 2) Display via FieldCare P01-xxxxxxxx-19-xx-xx-en-152 Endress+Hauser...
  • Page 228 PCB COUNT:T > Tmax TOTALIZER 2 PCB MAX. TEMP. TOTAL. 2 OVERFLOW PCB COUNT: T< Tmin PCB MIN. TEMP. RESET PEAKHOLD There are parameters that are only displayed if other parameters are appropriately configured. These parameters are indicated with a "*". P01-xxxxxxxx-19-xx-xx-en-170 Endress+Hauser...
  • Page 229 Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus Appendix Endress+Hauser...
  • Page 230 Pressure Flow Level Tank content SIM. PRESSURE SIM. FLOW VALUE SIM. TANK CONT. SIM. ERROR NO. SIM. LEVEL There are parameters that are only displayed if other parameters are appropriately configured. These parameters are indicated with a "*". P01-xxxxxxxx-19-xx-xx-en-154 Endress+Hauser...
  • Page 231: Index

    Alternate Data........88 T > Tmax ........78 Endress+Hauser...
  • Page 232 ELECTRONIC_OVER_TEMPERATURE ....78 CUSTOMER_UNIT_FLOW ......84 Endress+Hauser...
  • Page 233 HYDR. PRESS MAX. (774)......168 FULL CALIB. (004), "Level easy height" level selection . . 153 Endress+Hauser...
  • Page 234 LO_ALM_OUT_D......102 MEASURED_VALUE_TREND ..... 79 Endress+Hauser...
  • Page 235 Pmax ALARM WINDOW (333)....208 Pmax PROC. CONN. (570) ..... . . 197 Endress+Hauser...
  • Page 236 Tmin Sensor ........76 SIM. PRESSURE (414)......206 Endress+Hauser...
  • Page 237 XD_ERROR........52 XD_SCALE ........95 Endress+Hauser...
  • Page 238 www.endress.com/worldwide BA00303P/00/EN/13.11 71139808 CCS/FM+SGML6.0 71139808...