Analog Window Triggering - National Instruments NI 6221 User Manual

Daq m series
Hide thumbs Also See for NI 6221:
Table of Contents

Advertisement

Chapter 11
Triggering
Hysteresis
Then signal must go below low threshold
before Analog Comparison Event asserts
Analog Comparison Event

Analog Window Triggering

Analog Comparison Event
M Series User Manual
First signal must go
above high threshold
Figure 11-6. Analog Edge Triggering with Hysteresis Falling Slope Example
An analog window trigger occurs when an analog signal either passes into
(enters) or passes out of (leaves) a window defined by two voltage levels.
Specify the levels by setting the window Top value and the window Bottom
value.
Figure 11-7 demonstrates a trigger that asserts when the signal enters the
window.
Top
Bottom
Figure 11-7. Analog Window Triggering Mode (Entering Window)
11-6
High threshold
(Level + Hysteresis)
Low threshold
(Level)