Counter n Gate Signal
© National Instruments Corporation
Routing Counter n Source to an Output Terminal
You can route Counter n Source out to any PFI <0..15> or RTSI <0..7>
terminal. All PFIs are set to high-impedance at startup.
The Counter n Gate signal can perform many different operations
depending on the application including starting and stopping the counter,
and saving the counter contents.
Routing a Signal to Counter n Gate
Each counter has independent input selectors for the Counter n Gate signal.
Any of the following signals can be routed to the Counter n Gate input.
•
RTSI <0..7>
•
PFI <0..15>
•
ai/ReferenceTrigger
•
ai/StartTrigger
•
ai/SampleClock
•
ai/ConvertClock
•
ao/SampleClock
•
di/SampleClock
•
do/SampleClock
•
PXI_STAR
•
Change Detection Event
•
Analog Comparison Event
In addition, Counter 1 Internal Output or Counter 1 Source can be routed to
Counter 0 Gate. Counter 0 Internal Output or Counter 0 Source can be
routed to Counter 1 Gate.
Some of these options may not be available in some driver software.
Routing Counter n Gate to an Output Terminal
You can route Counter n Gate out to any PFI <0..15> or RTSI <0..7>
terminal. All PFIs are set to high-impedance at startup.
7-27
Chapter 7
Counters
M Series User Manual