Chapter 2 Initial Operation
Output Checkout

Current Output Checkout

The following steps check basic current functions with a short across the power
supply's output.
Power
1 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled (the
OFF
** annunciator turns on); and the knob is selected for voltage control.
25V
2 Connect a short across (+) and (-) output terminals with an insulated test lead.
Use a wire size sufficient to handle the maximum current (See "Table 7-1 Wire Rating"
on page 153 in chapter 7).
Output
On/Off
3 Enable the output.
The
are lit. The
The blinking digit can be adjusted by turning the knob. Notice that the display is in
the meter mode. ''Meter mode'' means that the display shows the actual output voltage
and current.
Display
Limit
4 Adjust the voltage limit value to 1.0 volt.
Set the display to the limit mode (the
voltage limit to 1.0 volt to assure CC operation. The
go back to normal mode, press the
several seconds .
Voltage
Current
5 Set the knob to the current control to check that the front-panel ammeter
properly responds to knob control.
Turn the knob clockwise or counter clockwise when the display is in the meter mode
(the
the voltmeter indicates nearly zero (the voltmeter will show the voltage drop caused
by the test lead).
*For Agilent E3633A Model
18
annunciator turns on); the 8V/20A* or 25V/7A** range is selected (the
annunciator turns off and the
OFF
or
annunciator turns on depending on the resistance of the test lead.
CV
CC
annunciator is off). Check that the ammeter responds to knob control and
Limit
**For Agilent E3634A Model
* or
**,
, and
8V
25V
OVP
annunciator will be blinking). Adjust the
Limit
annunciator will turn on. To
CC
Display
key again or let the display time out for
Limit
* or
8V
annunciators
OCP

Hide quick links: