Error Messages

Errors are retrieved in first-in-first-out (FIFO) order. The first error returned is the
first error that was stored. Errors are cleared as you read them over the remote
interface. When you have read all errors from the queue, the
turns off and the errors are cleared. The power supply beeps once each time an error
is generated.
If more than 20 errors have occurred, the last error stored in the queue (the most
recent error) is replaced with -350, ''Too many errors''. No additional errors are stored
until you remove errors from the queue. If no errors have occurred when you read the
error queue, the supply responds with +0, ''No error'' over the remote interface or
''NO ERRORS'' from the front panel.
The error queue is cleared by the *CLS (clear status) command or when power is
cycled. The errors are also cleared when you read the queue.
The *RST (reset command) command does not clear the error queue.
• Front-panel operation:
Store
Press
(
) key to return to front-panel operation mode if the power supply
Local
Local
is in remote operation mode.
3: ERR -102
If the
annunciator is on, press
ERROR
scroll through the error numbers. Press
<
>
Press
or
key to decrease or increase the scrolling speed of the text on the
display. All errors are cleared when you exit the menu by pressing the
let the display time out for about 30 seconds.
• Remote interface operation:
SYSTem:ERRor?
Errors have th5e following format (the error string may contain up to 80 characters).
-102, ''Syntax error''
112
Error
key to view the errors. Use the knob to
>
to view the text of the error message.
Read and clear one error from the error queue
annunciator
ERROR