Agilent Technologies E8362B Service Manual page 93

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Isolate the remaining three assemblies:
• remove the A15 CPU board
• remove the A4 power supply
• This leaves only the A14 system motherboard installed. If the measurements are still
incorrect, this is the suspected faulty assembly.
Table 4-2
Power Supply Voltages and Resistances
Measurement
Location
A
B
C
D
E
F
G
H
Service Guide E8364-90026
Supply
Approximate
(Vdc)
Resistance (Ω)
+5.0
285
-15.0
12.7 k
-5.0
7.2 k
+9
4.0 k
+15
3.0 k
+22
2.8 k
+22
890
+32
2.6 k
Troubleshooting
Power Up Troubleshooting
4-11