Agilent Technologies E8362B Service Manual page 91

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Figure 4-3 E8356-60021 Synthesizer/Reference Extender Measurement Points
Table 4-1
Extender Board Measurement Points
Measurement
Location
Signal Description
A
B
C
D
E
F
G
H
1. DVM=Digital volt meter
If All Supply Voltages Are Present
If all of the supplies have measured within tolerances, and the instrument still is not
functioning properly, refer to
Service Guide E8364-90026
power supply
power supply
power supply
power supply
power supply
power supply
power supply
power supply
"Rear Panel Troubleshooting" on page
Test Equipment
1
Used
DVM
DVM
DVM
DVM
DVM
DVM
DVM
DVM
Troubleshooting
Power Up Troubleshooting
Expected
Level (Vdc)
+5.0
−15.0
−5.0
+9
+15
+15
+22
+32
4-22.
4-9