PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
System Verification Procedure
1. Connect a cable to the analyzer as shown in
outputs, connect a printer to the analyzer. For the printer, ensure that the correct driver
is loaded and the printer is defined as the default printer. Refer to the embedded help in
the analyzer for printer setup. Let the analyzer warm up for at least 90 minutes.
Figure 3-11 System Verification Test Setup
2. Insert the verification kit disk into the analyzer disk drive.
3. On the System menu, point to Service, and then click System Verification. The System
Verification dialog box is displayed; refer to
Figure 3-12 System Verification Dialog Box
Service Guide E8364-90026
Figure
3-11. If you desire printed test
Figure
3-12.
Tests and Adjustments
System Verification
3-25