Agilent Technologies E8362B Service Manual page 137

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
3. Connect the spectrum analyzer to A10J11.
4. The spectrum analyzer should measure a signal at 5 MHz.
5. If no 5 MHz signal is present, replace the A10 frequency reference board. Refer to
"Removing and Replacing the A6, A8, A9, and A10 Boards" on page
6. If the 5 MHz signal is present, reconnect cable W115, and then continue with
the A9 Fractional-N Synthesizer Band 0 and 1 Output" on page
Checking the A9 Fractional-N Synthesizer Band 0 and 1 Output
Perform this procedure if you observe a problem in band 1 of the frequency offset receiver.
1. Refer to the block diagram at the back of this chapter and to
Options Including 080" on page
fractional-N synthesizer board.
CAUTION
Be careful not to damage the center pins of the semi-rigid cable. Some flexing
of the cables is necessary to measure the output.
2. Disconnect W92 from A9J101.
3. Connect the spectrum analyzer to A9J101.
4. Set the network analyzer for a 500 MHz CW frequency and observe the spectrum
analyzer measurement.
• The spectrum analyzer should measure a signal at 508.33 MHz.
5. If the 508.33 MHz signal is not present and the 5 MHz reference signal is present from
"Checking the A10 5 MHz Reference Output, All Bands,"
synthesizer board. Refer to
on page
7-18.
6. If the signal is present, reconnect cable W92, and then continue with
Fractional-N Synthesizer Band 2 Output" on page
Checking the A9 Fractional-N Synthesizer Band 2 Output
Perform this procedure if you observe a problem in band 2 of the frequency offset receiver.
1. Refer to the block diagram at the back of this chapter and to
Options Including 080" on page
fractional-N synthesizer board.
CAUTION
Be careful not to damage the center pins of the semi-rigid cable. Some flexing
of the cables is necessary to measure the output.
2. Using a 5/16-inch torque wrench, disconnect W91 at A9J106.
3. Connect the spectrum analyzer to A9J106.
4. Set the network analyzer to a CW frequency of 1 GHz.
• The spectrum analyzer should measure a signal at 1.008 GHz.
Service Guide E8364-90026
6-18. Locate semi-rigid cable W92, at the A9
"Removing and Replacing the A6, A8, A9, and A10 Boards"
6-18. Locate semi-rigid cable W91, at the A9
Measurement System Troubleshooting
7-18.
4-55.
"Top Cables–All, All
replace the A9 fractional-N
"Checking the A9
4-55.
"Top Cables–All, All
Troubleshooting
"Checking
4-55