Agilent Technologies E8362B Service Manual page 44

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

Tests and Adjustments
Preliminary Checks
3. Click Begin.
4. If shorts and opens are not connected to all ports, you will be prompted to connect them
as they are needed.
5. The result of the operator's check will be shown as a PASS or FAIL next to each test
(refer to
Figure
3-3). The PNA Operator's Check dialog box will look different for
different PNA model numbers and installed options. Some of the tests are performed
only if the appropriate options are installed in the PNA.
Figure 3-3
Operator's Check Dialog Box
If the Operator's Check Fails
1. Clean the test ports, shorts, and adapters. Torque connections to specification. Repeat
the check.
2. If the check still fails, suspect a faulty component. Refer to
Troubleshooting" on page 4-29
component.
3-12
to begin troubleshooting to determine the faulty
PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
"Measurement System
Service Guide E8364-90026