Agilent Technologies E8362B Service Manual page 131

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Checking Port 2 Power (B Signal)
The object of this check is to verify the power of the B signal across the entire frequency
range. Perform this test if there is an observed problem only with the channel B trace. The
ten second sweep is slow enough to allow you to observe the output power on the power
meter as the sweep occurs.
1. Connect the power sensor to Port 2.
2. Observe the power reading on the power meter as the sweep occurs on the analyzer.
3. The measured output power on the power meter should be within −12 dBm ±4 dB for
E836xB with no options or −17 dBm ±4 dB for E836xB with options, over the entire
frequency range.
Checking the Signal through the Signal Separation Path
For all of the following checks, refer to the block diagrams at the end of this chapter and to:
"Bottom Assemblies, No Options" on page 6-20
"Bottom Assemblies, Option UNL" on page 6-24
Options UNL/014/016/080/081" on page 6-58
Port 1 trace loss in the signal separation group is due to one or more of the following
assemblies being defective:
• A22 switch splitter
• A25 test port 1 coupler
• A36 60-dB step attenuator (Option UNL)
• A38 bias tee (Option UNL)
Port 2 trace loss in the signal separation group is due to one or more of the following
assemblies being defective:
• A22 switch splitter
• A26 test port 2 coupler
• A37 60-dB step attenuator (Option UNL)
• A39 bias tee (Option UNL)
Equipment Used for These Tests
Equipment Type
Spectrum analyzer
1. Must be capable of measuring a signal at 1 GHz.
To determine which assembly is defective, check the signal at each available measurement
point in the signal path from the A22 switch splitter to the test port coupler.
For Port 1 measurements, set the network analyzer for an S
frequency of 1GHz.
Service Guide E8364-90026
and
thru
"Bottom Semi-Rigid Cables,
(as required)
Models
Model or
Used With
Part Number
All
8565E
Troubleshooting
Measurement System Troubleshooting
Alternate Model
or Part Number
1
856xE
measurement with a CW
11
4-49

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363bE8364b

Table of Contents