PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Data are listed here as a convenience only. Detailed instrument specifications
IMPORTANT
are listed in the embedded help in the network analyzer.
Table 8-1
Error Term Data
Parameter
Directivity
Source Match
Standard
Option UNL, 014,
or UNL & 014
Load Match
Standard
Option UNL, 014,
or UNL & 014
b
Reflection Tracking
Transmission
b
c
Tracking
,
Crosstalk (Isolation)
Standard
Option UNL or 014
Option UNL & 014
b
Option 080 enabled
a. Uncorrected system performance. The values in the table apply over an environmental
temperature of 23 °C ±3 °C, with less than 1 °C deviation from the calibration
temperature.
b. Typical performance.
c. Transmission tracking performance is strongly dependent on cable used. These typical
specifications are set based on the use of Agilent through cable part number 85133-60016.
d. 0 Hz offset.
e. The value is -85 dB for frequencies below 1 GHz.
f. The value is -95 dB from 45 MHz to 1 GHz.
If Error Terms Seem Worse than Expected
To verify that the system still conforms to specifications, perform a system verification.
Refer to
Chapter 3, "Tests and
Service Guide E8364-90026
a
Frequency Range
10 MHz
45 MHz
to
to
b
2 GHz
45 MHz
>22 dB
>24 dB
>17 dB
>23 dB
>12 db
>18 dB
>20 dB
>23 dB
>12 dB
>17 dB
±1.5 dB
±1.5 dB
±2.0 dB
±1.5 dB
-85 dB
e
-100 dB
-85 dB
e
-100 dB
-85 dB
e
-98 dB
-85 dB
d
,
f
-108 dB
Adjustments".
General Purpose Maintenance Procedures
2 GHz
10 GHz
20 GHz
to
to
to
10 GHz
20 GHz
40 GHz
>22 dB
>16 dB
>16 dB
>16 dB
>14 dB
>10 dB
>14 dB
>12 dB
>8 dB
>14 dB
>10 dB
>9 dB
>13 dB
>10 dB
>9 dB
±1.5 dB
±1.5 dB
±1.5 dB
±2.0 dB
±2.5 dB
±3.5 dB
-110 dB
-110 dB
-108 dB
-109 dB
-109 dB
-106 dB
-108 dB
-107 dB
-104 dB
-115 dB
-120 dB
-115 dB
Error Terms
40 GHz
45 GHz
to
to
45 GHz
50 GHz
>15 dB
>13 dB
>9 dB
>5.5 dB
>7 dB
>4 dB
>9 dB
>7 dB
>8 dB
>6 dB
±2.0 dB
±2.0 dB
±4.0 dB
±4.5 dB
-105 dB
-100 dB
-103 dB
-98 dB
-100 dB
-95 dB
-112 dB
-105 dB
8-11