E8362B, E8363B, E8364B
Checking the A18 MA 20 Output
Perform this procedure if you observe a problem in any band and all signals from the A8
fractional-N synthesizer are present.
1. Refer to the block diagram at the back of this chapter and to
Cables, No Options" on page
CAUTION
Be careful not to damage the center pins of the semirigid cable. Some flexing
of the cables is necessary to measure the output.
2. Using a 5/16-inch torque wrench, disconnect W17 at the A18 MA20.
3. Connect the spectrum analyzer to the open A18 connector.
4. For each frequency band in which problems were seen: Set the network analyzer to each
of the frequencies indicated in the
column in
Table 7 on page
5. Set the spectrum analyzer to measure each of the corresponding frequencies in the
column labeled:
"Expected A18 Output Frequency (GHz)"
6. If the signals are <−30 dBm or not present, replace the A18 MA 20. Refer to
and Replacing the A18 MA 20" on page
7. If the signals are present, reconnect cable W17, and then continue with
A20 LODA Output" on page
Checking the A20 LODA Output
Perform this test if you observe a problem in all bands or a single channel.
1. Refer to the block diagram at the back of this chapter and to
Cables, No Options" on page
LODA.
2. Using a 5/16-inch torque wrench, disconnect the cable associated with the faulty
channel (W11, W12, W13, or W14) and connect the spectrum analyzer to the A20 LODA
connector.
3. Set the network analyzer to any one of the frequencies listed in
4. Set the spectrum analyzer to measure the corresponding A18 output frequency listed in
Table
7.
5. If the signal is not present, replace the A20 LODA. Refer to
the A20 LODA" on page
6. If the signal is present, reconnect the cable, and then go to
Group" on page
4-52.
Service Guide E8364-90026
6-22. Locate semirigid cable W17, at the A18 MA 20.
"Network Analyzer Source Frequency Setting (GHz)"
4-46.
7-34.
4-47.
6-22. Locate cables W11, W12, W13, and W14, at the A20
7-38.
Troubleshooting
Measurement System Troubleshooting
"Bottom Semi-Rigid
in
Table 7 on page
4-46.
"Removing
"Checking the
"Bottom Semi-Rigid
Table 7 on page
"Removing and Replacing
"Checking the Receiver
4-46.
4-47