3. Point to Programs, Accessories, and then open Windows Explorer.
4. Navigate to C:\Program Files\Agilent\Network Analyzer.
5. Copy each of the backup files from the floppy disk or USB memory drive to the HDD.
6. Navigate to D:\CalFiles\.
7. Copy the backup files prefixed with mxcalfile_ from the floppy disk or USB memory
drive to the HDD.
8. After all files have been copied, remove the storage device containing the backup files:
either the floppy disk from the analyzer's disk drive, or the USB memory drive from the
USB port.
9. Exit Windows Explorer .
Post-Repair Procedures
1. Adhere the new Certificate of Authenticity (license) label to the outer cover in the
location specified in
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
Service Guide E8364-90026
Figure 7-1 on page
7-7.
Table 7-2 on page
Repair and Replacement Procedures
Removing and Replacing the A41 Hard Disk Drive (HDD)
7-80.
7-67