Tests and Adjustments
Preliminary Checks
Cable Connector Repeatability Check
NOTE
The connector repeatability measurement should be done at the test port as
well as at the end of the test port cable.
1. With the test port cable still connected to Port 1, connect a broadband load to the other
end of the cable.
2. Press Preset.
3. On the Channel menu, click Average. In the Average dialog box, click the Average ON
check box. In the Average Factor box, type 100 or click the arrows to select 100. Click
OK.
4. Wait for the analyzer to average the measurement 100 times (approximately five
seconds).
5. To normalize the data trace:
a. On the Trace menu, click Math/Memory.
b. In the Math/Memory dialog box, click the Data->Memory button.
c. In the Data Math list, select Data/Memory.
d. Under Trace View Options, make sure Data Trace is selected.
e. Click OK
6.
To adjust the display scale:
a. On the Scale menu, click Scale.
b. In the Scale Per Division box, click the arrow to select 0.5 dB.
c. In the Level box under Reference click the arrow to select 0 dB.
d. Click OK.
7. Disconnect and then reconnect the cable to the test port. Tighten the connection to the
specified torque for the connector type.
8. On the Channel menu, click Restart Avg.
9. Look at the trace for spikes or modes.
10. To re-normalize the data trace of the reconnected cable:
a. On the Trace menu, click Math/Memory.
b. In the Math/Memory dialog box, click the Data->Memory button.
c. Click OK.
11. Repeat steps 7 through 9 at least three times to look for modes. Modes appear when a
harmonic of the source fundamental frequency is able to propagate through the cable or
connector. It is helpful to print a plot of the trace each time to compare several
connections. If any mode appears each time the cable is connected and reconnected,
measurement integrity will be affected.
For a typical response of cable connector repeatability, see
3-18
PNA Series Microwave Network Analyzers
E8362B, E8363B, E8364B
Figure
3-7.
Service Guide E8364-90026