Agilent Technologies E8362B Service Manual page 140

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

Troubleshooting
Measurement System Troubleshooting
• The spectrum analyzer should measure a signal at 1 GHz.
5. If the measured signal is not present, replace A46 switch. Refer to
Replacing the A45 and A46 Reference Channel Transfer Switches (Option 081)" on
page
7-70.
6. If the measured signal is present, reconnect W60.
7. From the
Channel
. Click
.
loop
OK
8. Remove semirigid cable W96 from A45 switch input.
9. Connect the spectrum analyzer to the W96 cable.
10.The spectrum analyzer should measure a signal at 1 GHz.
11. If the measured signal is not present replace, A46 switch. Refer to
Replacing the A45 and A46 Reference Channel Transfer Switches (Option 081)" on
page
7-70.
12. If the measured signal is present then reconnect W96.
13. Remove W99 from the A45 switch output.
14. Connect the spectrum analyzer to the output of the A45 switch.
15. The spectrum analyzer should measure a signal at 1 GHz.
16. If the measured signal is not present replace A45 switch. Refer to
Replacing the A45 and A46 Reference Channel Transfer Switches (Option 081)" on
page
7-70.
17. If the instrument passes all the tests to this point but you still have a problem with
your PNA, then refer to
4-58
menu, click
. In the
Test Set
"Contacting Agilent" on page
PNA Series Microwave Network Analyzers
dialog box, click
Test Set
2-13.
Service Guide E8364-90026
E8362B, E8363B, E8364B
"Removing and
Internal: bypass R1
"Removing and
"Removing and