NORMA 4000/5000
Operators Manual
View Details of a Measured Value
Set Frequency Range
The default frequency range is set to a maximum of half the sampling frequency.
8-22
Shop for Power Metering products online at:
To adjust the scale, use the cursor keys:
Cursor Key
Left or right
Up or down
Enter
Esc
3. Adjust the scale of the scale of the axis, using the cursor keys, and press Enter or
Esc.
The graph with the adjusted axis is shown.
1. Press function key Detail.
The details of measured value U1 (voltage) are displayed.
2. Press function key Detail again.
The details of measured value I1 (current) are displayed.
3. Press function key Detail again.
The details of measured value P1 (power) are displayed.
4. Press function key Detail again.
The details of measured value U12 (phase-to-phase voltage) are displayed.
5. To return to the overview of measured values of the selected channel, press
function key Detail again (or Esc from any of the previous detail screens).
For signals with a lower frequency (for example, 10 Hz), the frequency
range must be adjusted; otherwise, the measurements would be inaccurate.