Configure Input Range; Automatic Range Adjustment (Auto); Manual Range Adjustment (Range); Configure Scale - Fluke NORMA 4000 Operator's Manual

Power
Hide thumbs Also See for NORMA 4000:
Table of Contents

Advertisement

Configure Input Range

You have the option to select automatic range configuration for the connected current
source (Auto). Alternatively, you can configure the range manually (Range).
With automatic configuration, the Power Analyzer determines and selects the correct
range for the connected current source.

Automatic Range Adjustment (Auto)

First field in column Auto is highlighted.

Manual Range Adjustment (Range)

To manually configure the range for I1, enter the range in amperes or, if shunts are used,
in volts.
First field in column Auto is highlighted.

Configure Scale

If you intend to use a shunt or a probe, you must adjust the scale for the output of the
measuring signal.
Shop for Power Metering products online at:
1. Press Enter.
2. Select ON and confirm by pressing Enter.
The selected settings are shown in column Auto.
3. If you wish to configure all three current channels in this way, press Set All.
All channels are now set to ON.
4. Save the configuration settings by pressing the SAVE function key.
1. Press Enter, select OFF and confirm by pressing Enter.
Automatic range adjustment is now disabled.
2. Move the cursor to the value in column Range and press Enter.
3. Select a value in amperes; if you use a shunt, select a value in volts.
When a value in volt is entered, automatic configuration (Auto) is set to Off.
Below Scale, option menu A/V is displayed.
1. Confirm by pressing Enter.
The settings are shown in column Range. Off is displayed in column Auto.
2. To configure all three current channels in this way, press Set All.
3. Save the configuration settings by pressing the SAVE function key.
The correct parameter settings are shown on the shunt or probe type plate.
www.
Configure Current and Voltage Channels
Note
Note
PowerMeterStore
Configuration
7-15
1.877.766.5412
.com
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Norma 5000

Table of Contents