Technical Data Fluke NORMA 4000/5000
General Technical Data
Compact system
Interface Commands
Housing
Weight
Dimensions (W,H,D)
Display
Operation
Mains connection
Measuring terminals
Calibration interval
Reference Conditions
Temperature
Humidity
Power supply
Power frequency
Warm up period
Ambient Conditions
Operating temperature range
Storage temperature range
Climatic class
Relative humidity
Altitude
Standards
Electrical safety
EN 61010-1/ 2. edition
EN 61558
EN 61010-2-031/61010-2-032
Electromagnetic compatibility
Emission
Immunity
Test voltages
Mains input housing (earth ground connector)
Mains connection measuring inputs
Measuring inputs - housing
Measuring inputs – measuring inputs
Shop for Power Metering products online at:
NORMA 4000
With 1 to 3 phases
Continuous averages
SCPI Vers.1999.0; legacy emulation of D5255 selectable
Protection Class 1 metal housing, IP 40
5 kg (11 lb)
237.0 mm (9.3 in.), 150.0 mm (3HU)
(5.9 in.), 315.0 mm (12.4 in.)
145 mm (5.7 in.), 320 x 240 pixel; background illumination and contrast
adjustable
Membrane keyboard, with cursor, function keys and direct functions
85 to 264 V AC (47 to 440 Hz) 120 to 300 V DC, Euro plug with switch
approximately 40 VA
4 mm safety sockets, 2 each / input; (for current inputsoptional binding post)
shunt connection via BNC socket
2 years
23 °C ±1 °C (71.6 °F to 75.2 °F)
< 60 % r.H.
115 V / 230 V ±10 %
50 Hz / 60 Hz
> 30 minutes
+5 to +35 °C (+41 °F to +95 °F)
−20 to +50 °C (−4 °F to +122 °F)
B2 (according to IEC 60654-1)
maximum 85 %, noncondensing
Below 2000 m
1000 V CAT II (600 V CAT III)
Degree of pollution 2, Protection class I
for transformer
for accessories
IEC 61326-1, class B
IEC 61326-1 / industrial locations
1.5 kV ac
5.4 kV ac
3.3 kV ac
5.4 kV ac
PowerMeterStore
www.
Technical Data
Technical Data Fluke NORMA 4000/5000
NORMA 5000
up to 6 phases
7 kg (15 lb)
447.0 mm (17.6 in.), 150.0 mm (3HU)
(5.9 in.), 315.0 mm (12.4 in.)
approximately 65 VA
.com
11
11-3