Incorrect Inlet Setup; Sensor Disabled - BW Technologies MicroDock II User Manual

Automatic test and calibration station
Hide thumbs Also See for MicroDock II:
Table of Contents

Advertisement

Incorrect Inlet Setup

If an inlet is not setup correctly during a calibration, the base station dis-
plays Fail and a hyphen (
S, CO, and LEL gases were not detected.
Example: H
2
If a test gas fails, verify that the
gas cylinder is not empty,
connections are attached correctly, and
inlets are setup correctly.
If the detector fails an audible or visual test,
cable test (
or VIS). For information regarding solutions, refer to
AUD
Troubleshooting.
) beside the gas type.
displays beside the appli-

Sensor Disabled

To prevent personal injury and/or property damage, enable or
replace the sensor immediately. Refer to the corresponding
detector user manual for sensor replacement procedures.
If a sensor is disabled, results from the sensor do not display on the base
station LCD.
displays where the gas type typically displays on the
LCD.
If a sensor is disabled, it will not be recorded in the results log.
MicroDock II
a Warning
Note
Calibration
71