Inserting The Gasalertmicro 5/Pid/Ir - BW Technologies MicroDock II User Manual

Automatic test and calibration station
Hide thumbs Also See for MicroDock II:
Table of Contents

Advertisement

Table 9. Inserting the GasAlertMicro 5/PID/IR
Item
1
Docking module lid
2
Release tabs
3
Diffusion adapter
4
Detector bay
5
Connector pins (charging model only)
Charge status indicator
6
(charging model only)
7
Connector outlets
8
Diffusion adapter release tab
Tool used to push down on diffusion
9
adapter release tab
If the GasAlertMicro 5/PID/IR is fitted with a calibration cap or
pump auxiliary filter, it must be removed prior to inserting it into
the docking module.
1. Activate the detector and wait until it is in normal operation.
2. Ensure the alligator clip is closed and the ring is resting flat on
the detector to prevent disruptions with data transmission.
Description
Note
3. Press the two release tabs on the docking module and raise the
lid.
Important: If the GasAlertMicro 5/PID/IR detector is fitted with a
pump, the diffusion adapter must be removed from the docking
module lid.
Figure 8. Removing the Diffusion Adapter
Using a small flathead screwdriver, press the release tab down-
wards. Pull the adapter forward slightly and then lift upwards to
remove.
MicroDock II
Installation
17