Inserting The Gasalertquattro - BW Technologies MicroDock II User Manual

Automatic test and calibration station
Hide thumbs Also See for MicroDock II:
Table of Contents

Advertisement

MicroDock II
User Manual

Inserting the GasAlertQuattro

Infrared or intense ambient light (sun or halogen) may interfere
with the base station/detector communication.
Figure 12. Inserting the GasAlertQuattro
24
a Caution
Table 12. Inserting the GasAlertQuattro
Item
1
GasAlertQuattro
2
Charge status
3
Connector pins
4
Detector bay
5
Module lid
6
Release tabs
If the GasAlertQuattro is fitted with a calibration cap or auxiliary
filter, it must be removed prior to inserting the detector into the
docking module.
When the GasAlertQuattro is inserted into a docking module,
the detector may take up to 20 seconds before beginning
charging.
1. Activate the detector and wait until it is in normal operation.
2. Ensure the alligator clip is closed and the ring is resting flat on
the detector to prevent disruptions with data transmission.
3. Press the two release tabs on the docking module and raise the
lid.
4. Insert the bottom of the detector (LCD up) at a 30° angle into the
detector bay and then lower the top into place.
Description
Note