Changing Dip Switch Settings

only)
The gas type for bump tests is defined by setting the dip switch within the
GasAlertClip Extreme docking module.
Table 16. Dip Switch Settings
Gas Type
Sulphur dioxide (SO
)
2
Oxygen (O
)
2
Hydrogen sulfide (H
S)
2
Carbon monoxide (CO)
34
(GasAlertClip Extreme
Dip Switch Setting
To change the dipswitch setting on a GasAlertClip Extreme docking
module refer to Tables 15 and 16, Figures 16 and 17, and complete the
following:
1. Deactivate the base station.
2. Remove the three Phillips pan-head screws from the front and
the six Phillips flat-head screws from the back cover plate.
3. Gently separate the docking module from the base station or
other docking modules. Refer to
4. Configure the gas type by setting the dip switches to the desired
gas. Refer to
Table 16.
5. Reconnect the GasAlertClip Extreme docking module. Refer to
Figure 20.
Refer to
Table 13.
for torque specifications.
6. Fasten the bottom cover plate using the six Phillips flat-head
screws. Fasten the remaining three Phillips pan-head screws to
the top of the docking module.
Table 17. Changing Dip Switch Settings
Item
1
Bottom cover plate
2
Phillips flat-head screw (6)
3
Phillips pan-head screw (3)
4
Barbed fitting ports
5
Dip switch
Figure 19.
for gas type and dip switch settings.
Description